User Manual 19 pages · 3.4 MB · ZH

The TOM-600 is a static resistance tester that measures surface, volume, and grounding resistance, with a measurement range of 20 kΩ to 2.0 TΩ. It uses menu-based operation, shows readings on a two-line 16-character LCD, and selects 10 VDC or 100 VDC test voltage automatically or manually. It also measures ambient temperature and relative humidity and has a built-in timer.

Measured data can be stored in the instrument memory and viewed with the supplied KL Read Out software after connecting via mini USB. The manual covers panel operation, timer, sampling, voltage and display settings, test procedures, standard and optional electrodes such as ME250 and 222002, maintenance, warranty, test examples, and control software. This manual is written in Chinese.

Read online

Kleinwächter TOM-600 specifications

Resistance Range20kΩ ~ 2.0TΩ
Temperature Range0 ~ 60℃
Humidity Range10% ~ 90%
Battery Life12 hours

Source: Kleinwaechter TOM-600 Electrostatic Resistance Tester Operation Manual

All Kleinwächter TOM-600 documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What test voltages does the TOM-600 support?

The instrument supports automatic switching between 10V and 100V, or manual selection of fixed 10V or 100V.

From: Kleinwaechter TOM-600 Electrostatic Resistance Tester Operation Manual

Can the instrument measure temperature and humidity?

Yes, it simultaneously tests and records ambient temperature and relative humidity during measurement.

From: Kleinwaechter TOM-600 Electrostatic Resistance Tester Operation Manual

What is the resistance measurement range of the TOM-600?

The resistance measurement range is 20 kΩ to 2.0 TΩ.

From: Kleinwächter TOM-600 TERA Ohmmeter 600 Static Resistance Tester Operation Manual

How does the TOM-600 select test voltage in automatic mode?

It uses 10 VDC below 200 kΩ and 100 VDC at 200 kΩ and above.

From: Kleinwächter TOM-600 TERA Ohmmeter 600 Static Resistance Tester Operation Manual