Key facts
- Brand
- IST
- Model
- SMD 0805
- Category
- Temperature monitor
- Documents
- 2
- Languages
- EN
- Last added
- 2012-05-11
Start here

IST SMD 0805 + 1206 Platinum Nickel Thin-Film Chip Sensor Datasheet
The SMD 0805 + 1206 thin‑film chip sensor from IST is a platinum or nickel temperature sensor designed for high‑volume, automated production lines. It combines the accuracy and stability of Pt/Ni elements with cost‑effective, lead‑free construction suitable for pick‑and‑place assembly.
Key data include temperature ranges of –50 °C to +150 °C (1‑P, 2‑P) and –50 °C to +250 °C (3‑P, 4‑P), TCR values of 3850 ppm/K (Pt) and 6180 ppm/K (Ni), and compact package dimensions of 2.0 × 1.2 mm (0805) and 3.2 × 1.6 mm (1206). The sheet also lists soldering limits, contact options, response times, self‑heating, and long‑term drift specifications.
IST SMD 0805 specifications
| Nominal resistance options | 100, 5000 or 1000Q |
|---|---|
| Temperature range | -50 °C to +150 °C (1P, 2P); -50 °C to +250 °C (3P, 4P) |
| Tolerance classes (Pt) | DIN class A, DIN class B, 2× DIN class B |
| Tolerance classes (Ni) | DIN, ¥% DIN (IST cl. A) |
| Temperature coefficient (TCR) | Pt: 3850 ppm/K, Ni: 6180 ppm/K |
| Soldering temperature limit | 235 °C < 8 s |
| Long‑term stability (drift) | Pt: 0.04 % after 1000 h at 130 °C; Ni: 0.1 % after 1000 h at 130 °C |
| Response time in water (0.4 m/s) | 0.258 s (1206), 0.28 s (0805) |
| Response time in air (1 m/s) | 5.08 s (1206), 4.08 s (0805) |
| Measuring current | 0.5 mA (100Q), 0.4 mA (500Q), 0.3 mA (10000) |
| Self‑heating (water) | 40 mW/°C |
| Package dimensions (Pt) | 0805: 2.0 × 1.2 mm, 1206: 3.2 × 1.6 mm |
| Package dimensions (Ni) | 1206: 3.2 × 1.6 mm |
Source: IST SMD 0805 + 1206 Platinum Nickel Thin-Film Chip Sensor Datasheet
All IST SMD 0805 documents by type
Ratings, dimensions and ordering codes, for selection and design work.
Questions answered in these manuals
What temperature range does the IST SMD 0805 + 1206 sensor support?
It operates from –50 °C to +150 °C for the 1‑pin and 2‑pin versions, and up to +250 °C for the 3‑pin and 4‑pin versions.
From: IST SMD 0805 + 1206 Platinum Nickel Thin-Film Chip Sensor Datasheet
What are the temperature coefficient (TCR) values for the platinum and nickel versions?
The platinum version has a TCR of 3850 ppm/K, while the nickel version has a TCR of 6180 ppm/K.
From: IST SMD 0805 + 1206 Platinum Nickel Thin-Film Chip Sensor Datasheet
What measurement current is recommended for a 100 Ω sensor?
The recommended measurement current for a 100 Ω sensor is typically 1 mA maximum.
From: IST SMD 0805/1206 Platinum/Nickel Thin-Film Chip Sensor Handbook
How fast does the 1206 sensor respond in water?
In water, the 1206 sensor reaches 63 % of the temperature change in about 0.25 seconds.
From: IST SMD 0805/1206 Platinum/Nickel Thin-Film Chip Sensor Handbook