This document serves as the user manual for GeoSIG Ltd's GSR, GCR, and AS series instruments, which are seismographs used for recording seismic data. The manual applies to specific firmware versions and provides comprehensive guidance on safety, incoming inspection, and the physical description of the housing and internal components like batteries and circuit boards.

The guide details the installation process, including site selection, power supply considerations, and mounting requirements. It explains how to configure parameters, utilize various triggering methods (threshold, STA/LTA, time, software, and external), and manage data recording and intensity scales. Additionally, it covers troubleshooting procedures, maintenance schedules for batteries, and cybersecurity recommendations.

Read online

GeoSIG GSR-12 specifications

Main Battery TypeFIAMM FG20721, 12V 7.2Ah
Backup Battery TypeVarta CR 1/2 AA SLF
Main Battery Replacement Interval3 years from date of production
Backup Battery Replacement Interval5 years from date of production

Source: GeoSIG GSR/GCR/AS Series Seismograph User Manual

All GeoSIG GSR-12 documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What are the replacement intervals for the internal batteries?

The main battery should be replaced every 3 years, and the backup battery every 5 years from the date of production.

From: GeoSIG GSR/GCR/AS Series Seismograph User Manual

What precautions must be taken before opening the instrument housing?

Users must flip the main switch to OFF, disconnect the power supply, disconnect the main battery, and ensure the Active Charge LED is OFF.

From: GeoSIG GSR/GCR/AS Series Seismograph User Manual

What safety precautions are needed before removing covers?

Before removing covers, you must flip the main switch to OFF, disconnect power supply, disconnect the main battery, and ensure the green Active Charge LED is OFF.

From: GeoSIG GSR-12/16/18 GCR-12/16 AS-12/16 User Manual

What specific steps must be taken before removing any covers inside the instrument?

You must flip the main switch to OFF, disconnect power supply, disconnect the main battery, and ensure the green Active Charge LED is OFF.

From: GeoSIG GSR-12/16/18 GCR-12/16 AS-12/16 User Manual