Elcometer 510 Manuals

ManualLib has 3 documents for the Elcometer 510 (Analyzer): 2 product leaflets, 1 user manual. Every file can be read online or downloaded as a PDF for free.

Product Leaflet 2User Manual 1

Key facts

Model
510
Category
Analyzer
Industry
Industrial
Documents
3
Languages
EN · ES
Last added
2026-11-21

Start here

Elcometer 5100 Payne Permeability Cup Operating Instructions

Elcometer 5100 Payne Permeability Cup Operating Instructions

User Manual 8 pages · 178 KB · EN

The Elcometer 5100 Payne Permeability Cup is a simple instrument used to determine the vapour transmission rate of unsupported films of paint, varnish, plastic, paper, and other materials. Manufactured from anodised aluminium, it is available in three sizes to suit different film testing requirements. The device operates by clamping a film over the cup, weighing the assembly before and after a set period, and calculating the water-vapour transmission rate based on the mass difference.

Read online

Elcometer 510 specifications

Measurement range, 10 mm dolly100 MPa / 14400 psi
Measurement range, 14.2 mm dolly50 MPa / 7200 psi
Measurement range, 20 mm dolly25 MPa / 3600 psi
Measurement range, 50 mm dolly4 MPa / 580 psi
Accuracy±1% of full scale
Dolly diameters10, 14.2, 20 and 50 mm
Pull rates0.1-1.4 MPa/s (15-203 psi)
Maximum stored readings60,000
Maximum batches2,500
Ingress protectionIP64
Display2.4 in (6 cm) TFT
Battery lifeUp to 200 pulls per charge
Battery recharge timeLess than 300 minutes
PowerRechargeable batteries or AC mains (Model T only)
Data outputUSB and Bluetooth
Warranty1 year, extendable to 2 years
StandardsASTM C1583, ASTM D4541, ASTM D7234-12, AS/NZS 1580.408.5, BS 1881-207, DIN 1048-2, EN 1015-12, EN 12636, EN 13144, EN 1542, EN 24624, ISO 16276-1, ISO 4624, JIS K 5600 5-7, NF T30-606, NF T30-062

Source: Elcometer 510 Automatic Pull-Off Adhesion Gauge Product Leaflet

All Elcometer 510 documents by type

User Manual 1

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What is the measurement range with a 10 mm dolly?

The Elcometer 510 measures up to 100 MPa (14400 psi) with a 10 mm dolly.

From: Elcometer 510 Automatic Pull-Off Adhesion Gauge Product Leaflet

How many readings can the Elcometer 510 store?

It can store up to 60,000 readings.

From: Elcometer 510 Automatic Pull-Off Adhesion Gauge Product Leaflet

What substrates can the Elcometer 510 test?

The Elcometer 510 is suitable for use on metal, wood, concrete, and other substrates.

From: Elcometer 510 Automatic Pull-Off Adhesion Tester Product Leaflet

What is the measurement accuracy of the Elcometer 510?

The instrument provides an accuracy of ±1% of full scale.

From: Elcometer 510 Automatic Pull-Off Adhesion Tester Product Leaflet

What standards can the Elcometer 5100 be used with?

The instrument can be used in accordance with ASTM D1653, ASTM E96, and ISO 7783.

From: Elcometer 5100 Payne Permeability Cup Operating Instructions

How is the water-vapour transmission rate calculated?

The rate is calculated using the formula P = 240 x Dm / A, where Dm is the rate of change in mass in mg/h and A is the area in square centimetres.

From: Elcometer 5100 Payne Permeability Cup Operating Instructions