Key facts
- Brand
- Applent
- Model
- AT820
- Category
- Other Electronics
- Industry
- Electronics
- Documents
- 2
- Languages
- ZH
- Last added
- 2026-11-16
Start here

Applent AT820 LCR Meter User Manual
The Applent AT820 is a general-purpose digital LCR bridge controlled by a high-performance microprocessor. It automatically measures inductance (L), capacitance (C), resistance (R), impedance (Z), quality factor (Q), and dissipation factor (D). The instrument features a 5-level sorting function with adjustable sound alerts and supports both series and parallel equivalent circuit modes. This manual is written in Chinese.
Read onlineApplent AT820 specifications
| Measurement Parameters | L, C, R, Z, D, Q |
|---|---|
| Test Frequency | 100Hz, 120Hz, 1kHz |
| Frequency Accuracy | ±0.02% |
| Test Level | 0.3Vrms |
| Level Accuracy | ±10% |
| Test Speed | 2.5 times/second |
| Basic Accuracy | 0.25% |
| Equivalent Mode | Series and Parallel |
| Measurement Terminals | 5-terminal |
| Power Supply Voltage | 198-252VAC |
| Power Supply Frequency | 47.5-52.5Hz |
| Power Consumption | Max 15VA |
| Fuse Rating | 250V, 0.5A Slow Blow |
| Operating Temperature | 0°C~55°C |
| Operating Humidity | <95%RH at 40°C |
All Applent AT820 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What are the measurement parameters of the AT820?
The instrument measures inductance (L), capacitance (C), resistance (R), impedance (Z), quality factor (Q), and dissipation factor (D).
What is the test frequency of the AT820?
The test frequencies are 100Hz, 120Hz, and 1kHz.
What should I do if the AT820 shows abnormal noise, smoke, or high temperature?
Stop operation immediately, disconnect the power cord, and contact Applent sales for repair to avoid fire or electric shock hazards.
What warranty does Applent provide for the AT820?
Applent offers a 90‑day exchange period and a three‑year free warranty, after which repairs are charged.