A&D MS-70 / MX-50 / MF-50 / ML-50 Analyzer User Manual
This manual is for the A&D MS-70, MX-50, MF-50 and ML-50 analyzers. It describes safety and compliance, precautions, installation, measurement procedures, programs, checks, printer and computer connections, data memory, calibration, functions, maintenance, specifications and index.
Safety sections warn about hot parts, flammable materials, electrical hazards and proper ventilation. Operating sections cover standard and quick modes, program storage, drying programs, self-check, temperature calibration for MS-70 and MX-50, and troubleshooting.
A&D MF-50 specifications
| Ambient operating condition | 5°C to 40°C (41°F to 104°F), 85%RH or less (no condensation) |
|---|---|
| Halogen lamp replacement recommendation | Replace after more than 5000 hours of use |
| Automatic temperature regulation | Maximum temperature is regulated to 160°C after one hour |
| High drying temperature behavior | At 200°C, the halogen lamp thermostat may work after 30 minutes |
Source: A&D MS-70 / MX-50 / MF-50 / ML-50 Analyzer User Manual
All A&D MF-50 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What ambient conditions are required for operation?
The analyzer should be operated at 5°C to 40°C (41°F to 104°F) and 85%RH or less with no condensation.
From: A&D MS-70 / MX-50 / MF-50 / ML-50 Analyzer User Manual
When should the halogen lamp be replaced?
A&D recommends replacing the halogen lamp when it has been used beyond 5000 hours.
From: A&D MS-70 / MX-50 / MF-50 / ML-50 Analyzer User Manual
What measuring principle does the moisture analyzer use?
It is based on the thermogravimetric principle: using the halogen lamp, moisture (%) and other results are obtained from the difference between the dried weight and the wet weight.
From: A&D MS-70/MX-50/MF-50/ML-50 Moisture Analyzer Instruction Manual
Which measurement modes are available?
There are 5 user modes: standard, quick, auto, timer and manual, plus drying programs (standard, gradual, step and quick drying; ML-50 uses only standard and quick drying).
From: A&D MS-70/MX-50/MF-50/ML-50 Moisture Analyzer Instruction Manual
What ambient conditions are required to operate the analyzer?
It should be operated at 5°C to 40°C (41°F to 104°F) and 85%RH or less with no condensation.
When should the halogen lamp be replaced?
Replacement is recommended when the lamp has been used beyond 5000 hours.