User manual for WITec alpha300/alpha500/alpha700 atomic force microscopes describing contact‑mode principles, setup, and operation.
This manual explains the principles of atomic force microscopy (AFM) in contact mode and provides detailed instructions for configuring the WITec Control software. It covers beam‑path alignment, cantilever selection, and the software interface used for AFM contact measurements.
The document guides users through the adjustment and alignment procedures, describes the speed‑button functions, control window layout, and data acquisition sequencers (Image Scan, Line Scan, Distance Curve), and includes recommended parameter settings for successful measurements.
| Cantilever length | 100–200 µm |
|---|---|
| Tip size | <10 nm |
| Recommended cantilever spring constant | <1 N/m |
| Proportional gain (recommended) | 5.0 |
| Integral gain (recommended) | 5.0 |
| Typical image scan points per line | 256–512 |
| Typical image scan lines per image | 256–512 |
| Typical image scan width/height | 2–30 µm |
| Typical time per line | 0.5–2 s |
Cantilevers with spring constants below 1 N/m are recommended for contact‑mode measurements.
Press the Auto‑Approach speed button or select the Start Approach button from the Microscope Z Control menu.