Operating manual for the Wide Plus 361 series smart single crystal transmitter, covering installation, calibration, specifications, and HART communication.
This document is the operating manual for the Wide Plus 361 series smart single crystal transmitters. It details the working principles, technical specifications, and installation requirements for measuring liquids, gases, and steam in industrial applications. The device provides (4~20) mA analog output and supports digital communication via the HART protocol. This manual is written in Chinese.
| Output Signal | (4~20) mA |
|---|---|
| Power Supply | 14.5~45V DC |
| Operating Temperature | -20℃~+70℃ |
| Storage Temperature | -46℃~110℃ |
| Process Temperature (Silicone Oil) | -40 to 121℃ |
| Electrical Interface | ANSI NPT1/2(F) or ISO M20×1.5 |
The amplifier component damping time constant is adjustable within the 0~100 second range.
Press and hold the Zero (Z) and Span (S) keys for more than 3 seconds until 'OPEN' appears on the LCD.
HART regulations allow a maximum of 15 intelligent instruments to be linked in series.