User manual for the FN Drop Value Determination Instrument by Top Instruments, covering safety, setup, operation and specifications.
The FN Drop Value Determination Instrument is designed to measure α‑amylase activity in grain samples by recording the time a stirring rod falls a set distance in a heated slurry. The instrument features a 6‑digit 8‑segment display, heating control, and automatic fault monitoring.
This manual provides safety requirements, unpacking checks, technical specifications, operating procedures, display explanations, and maintenance instructions. It also includes calibration methods for the water‑bath temperature and contact information for Top Instruments. This manual is written in Chinese.
| Working voltage | AC 220 V ±10 % 50 Hz |
|---|---|
| Repeatability | ≤ 10 % of average value |
| Stirring rod mass | 25 ± 0.5 g |
| Water‑bath heating power | 600 W |
| Viscosity tube inner diameter | 21 ± 0.2 mm |
| Viscosity tube outer diameter | 23.8 ± 0.25 mm |
| Viscosity tube inner wall height | 220 ± 0.3 mm |
| Instrument dimensions | 420 × 340 × 350 mm |
| Instrument weight | 25 kg |
It shows that the heating element is on and the water‑bath temperature has reached the set point, so the instrument is ready for the measurement.