TSKgel SuperAW Series Packed Columns for Polar Organic Solvent Size Exclusion Chromatography

Instruction manual for TSKgel SuperAW Series packed columns used in size exclusion chromatography with aqueous and polar organic solvent systems.


TSKgel SuperAW Series Packed Columns for Polar Organic Solvent Size Exclusion Chromatography - cover page
Brand
TSKgel
Category
Health Care
Document type
User Manual
Language
English
Pages
18
File format
PDF
File size
659 KB
Published
03 November, 2026
Updated
03 November, 2026
MD5 checksum
185BA0B64A9D32E7FC0F5E90A985513D

About this manual

This manual provides instructions for unpacking, installing, storing, and using TSKgel SuperAW Series packed columns for size exclusion chromatography. It covers solvent selection, flow rate, temperature, sample preparation, and troubleshooting. Safety precautions are emphasized, including handling flammable solvents and preventing bubble entry into the column.

The manual includes sections on column parts, installation procedures, storage conditions, solvent replacement, degassing, and column maintenance. It also provides guidance on connecting multiple columns and preventing performance degradation due to pulsation or improper solvent use.

What's inside

  • Introduction
  • Unpacking
  • Column Parts
  • Installation
  • Storage of Column
  • Selection of Solvents
  • Flow Rate
  • Temperature for Usage
  • Preparation of Sample Solutions
  • Measurement of Number of Theoretical Plates and Asymmetry Factor

Specifications

Column TypeSolvent-resistant, solvent-exchangeable, high-performance and semi-micro SEC-packed columns
Solvent CompatibilityAqueous solvent systems and various polar organic solvent systems (mixtures of water and organic solvents)
Flow Rate for Rinsing6.0 mL/min for 5.0 mm inner diameter, 15 cm length column

Frequently asked questions

What solvents are compatible with the TSKgel SuperAW Series columns?

The columns are compatible with aqueous solvent systems and various polar organic solvent systems (mixtures of water and organic solvents).

How should I handle solvent replacement in the SuperAW Series columns?

Replace the ion-exchanged water with the solvent to be used at the flow rate shown in Table 3. Avoid frequent solvent replacement to prevent column performance deterioration.

What precautions should I take to prevent bubble entry into the column?

Always install the column after removing bubbles from the entire piping system. Introducing bubbles can cause channeling and deteriorate column performance.

Download PDF (659 KB)