Teleflex Arrow EZ-IO Intraosseous Vascular Access System Pocket Guide

French pocket guide for the Teleflex Arrow EZ-IO intraosseous vascular access system, covering proximal humerus insertion, removal, and components.


Teleflex Arrow EZ-IO Intraosseous Vascular Access System Pocket Guide - cover page
Brand
Teleflex
Category
Professional medical equipment
Document type
User Manual
Language
French
Pages
2
File format
PDF
File size
2.5 MB
Published
21 November, 2026
Updated
21 November, 2026
MD5 checksum
270CB189255B3B8E9FBD9043468C1FC4

About this manual

This French pocket guide from Teleflex describes the Arrow EZ-IO intraosseous vascular access system for proximal humerus insertion in adults and children.

It covers insertion site identification, insertion and removal steps, system components such as the motorized insertion device and 15 G needles, and includes references and prescription-only warnings. This manual is written in French.

What's inside

  • Proximal humerus insertion site identification
  • Insertion and removal technique
  • Removal
  • EZ-IO system components
  • Growth plate
  • Additional considerations for the motorized insertion device
  • References

Specifications

Needle length45 mm
Needle length25 mm
Needle length15 mm
Needle gauge15 G
Black depth mark5 mm
Adult flush volume5-10 ml
Pediatric flush volume2-5 ml
Low battery level10%

Frequently asked questions

What does the black mark on the needle indicate?

The 5 mm black mark must be visible above the skin before insertion.

How much saline is used to flush the EZ-IO catheter?

Use 5-10 ml for adults or 2-5 ml for children.

What does a blinking red LED on the insertion device mean?

It means the device has only 10% battery life remaining and should be replaced.

Is the device available by prescription only?

Yes, the document states it is available by prescription only.