Instruction sheet for installing and operating the Teledyne Isco 6712Ci LTE cellular modem, covering APN setup, email alerts, SMS commands, and SIM card use.
The 6712Ci LTE cellular modem from Teledyne Isco attaches to the 6712 sampler and can transmit SMS text message alarms to cell phones and/or email addresses, and receive SMS text message commands via the internet. This instruction sheet covers programming the APN, activating and setting up the modem, power settings, notification lists, email setup, SMS commands, dual sampler setup, and SIM card installation.
The modem requires sampler firmware version 3.02 or newer and works only with C0 hardware; when the 6712Ci modem is used, the baud rate is locked at 9600 bps. The 6712Ci LTE Cellular Modem Kit (modem module with magnetic mount antenna) is part #60-9004-404, and the antenna alone is part #60-4804-035. A Micro SIM card is required for any LTE units, including the 6712Ci.
| Baud rate | Locked at 9600 bps |
|---|---|
| Required sampler firmware | 3.02 or newer (C0 hardware only) |
| Modem kit part number | 60-9004-404 |
| Antenna part number | 60-4804-035 |
| SIM card type | Micro SIM |
| Notification list | Up to 3 phone numbers and 3 email addresses |
Sampler firmware version 3.02 or newer is required, and it will only work with C0 hardware. You can verify the hardware version under VIEW REPORT > SYSTEM IDS.
Only Micro SIM cards work with the 6712Ci modems; a Micro SIM card is required for any LTE units, including the 6712Ci.
Send a text message to the 6712Ci modem's phone number containing the string "6712" plus a command such as START or STATUS. Commands are not case-sensitive and do not require spacing.
When the 6712Ci modem is used, the baud rate is locked at 9600 bps, so make sure the same setting is selected in Flowlink or Hyperterminal.