User manual for Shidai HLN-11 series Leeb hardness tester covering operation, maintenance, calibration, and troubleshooting.
This manual provides detailed instructions for operating, maintaining, and troubleshooting the Shidai HLN-11 series Leeb hardness tester. It covers instrument components, test principles, sample preparation, operation procedures, calibration using standard blocks, and repair services. The device is designed for industrial metal hardness testing and complies with Chinese national standards.
The HLN-11 series includes two models: HLN-11A (compatible with D-type impactors) and HLN-11C (compatible with D-type and various non-standard impactors). Both models operate similarly and support multiple hardness scales including HL, HRC, HRB, HB, HV, and HS. The instrument features automatic impactor identification, data logging, and a printer for outputting test results. This manual is written in Chinese.
| Measurement Range | HL, HRC, HRB, HB, HV, HS scales |
|---|---|
| Accuracy | Relative error ±0.8% (HLD=800) |
| Operating Temperature | 0~40℃ |
| Operating Voltage | 4.7V~6.0V |
| Weight | 0.675Kg (standard configuration) |
It complies with 《里氏硬度计技术条件》 ZB N71 010-90, 《里氏硬度计》检定规程 JJG 747-91, and 《金属里氏硬度试验方法》 GB/T 17394 -1998.
Use standard Leeb hardness blocks to calibrate the instrument according to JJG747-91, with specified error tolerances for each impactor type.