SH muffle furnace manual for SH-FU-3/5/7/14/27MG and SH-FU-3/5/7/11MGE: installation, temperature settings, specs, troubleshooting.
This manual covers the SH-FU-3/5/7/14/27MG and SH-FU-3/5/7/11MGE muffle furnaces. It lists key features such as digital temperature and time display, multi key-in sensor types, precise temperature control, timer-end alarm, over-current protection, auto backup, and, for the FU-xxMG series, over-temperature protection and a separated body and control box.
The document explains unpacking and installation, leveling the unit, connecting power, and setting the OVER TEMP value at least 30°C above the target. It then describes TEMP, TIME, and BEEP mode settings for FU-xxMG and FU-xxMGE series, temperature offset setting, specification tables, and troubleshooting.
| Maximum temperature (MG series) | 1200°C (2192°F) |
|---|---|
| Maximum temperature (MGE series) | 1050°C |
| Temperature controller | Digital PID Controller |
| Sensor | K type |
| Heater element (MG series) | KANTHAL A-1 with ARD molding |
| Heater element (MGE series) | KANTAL A-1 with CERAMIC FIBER |
| Chamber capacity (MG series) | 2.9 L, 4.5 L, 14 L, 27 L |
| Chamber capacity (MGE series) | 3 L, 5.1 L, 7.4 L, 11.2 L |
The MG series has a maximum temperature of 1200°C (2192°F).
It should be set at least 30°C higher than the target set temperature; for example, over 830°C when the controller is set to 800°C.
Press and hold the MODE button for 3 seconds.
00.00 means an infinite value, so the unit runs until the user manually turns it off.