SEB SEB083Hu Enzyme-linked Immunosorbent Assay Kit Instruction Manual

The SEB SEB083Hu ELISA kit manual provides step‑by‑step instructions for measuring human SOD2 with a sandwich immunoassay.


SEB SEB083Hu Enzyme-linked Immunosorbent Assay Kit Instruction Manual - cover page
Brand
SEB
Category
Health Care
Document type
User Manual
Language
English
Pages
8
File format
PDF
File size
658 KB
Published
30 September, 2026
Updated
30 September, 2026
MD5 checksum
3E86C8E9CC1E50F212914AF7391E9863

About this manual

This manual describes the intended use, reagents, storage conditions, sample handling, and detailed assay procedure for the SEB083Hu 96‑test ELISA kit that quantifies mitochondrial Superoxide Dismutase 2 (SOD2) in human biological samples. It also includes calculation methods, performance specifications, and troubleshooting guidance.

The kit contains a pre‑coated 96‑well strip plate, standards, detection reagents, substrate, stop solution, and wash buffer. Users must follow the described incubation times, washing steps, and reading conditions (450 ± 10 nm) to obtain reliable results.

What's inside

  • Intended Use
  • Reagents and Materials Provided
  • Materials Required but Not Supplied
  • Storage of the Kits
  • Sample Collection and Storage
  • Reagent Preparation
  • Assay Procedure
  • Test Principle
  • Calculation of Results
  • Precision and Sensitivity

Specifications

Detection range0.78‑50 ng/mL
Sensitivity (LLD)< 0.37 ng/mL
Recommended microplate readerBandwidth ≤ 10 nm, OD range 0‑3 at 450 ± 10 nm
Unopened kit storage temperature-20 °C for plate, 4 °C for other reagents
Opened kit usage windowUse remaining reagents within 1 month

Frequently asked questions

What is the intended use of the SEB083Hu kit?

The kit is a sandwich enzyme immunoassay for in‑vitro quantitative measurement of SOD2 in human serum, plasma, tissue homogenates, cell lysates and other biological fluids.

How should the kit be stored?

Unopened reagents such as the pre‑coated plate, standards and detection reagents should be stored at –20 °C, while other components are stored at 4 °C. After opening, keep reagents at the same temperatures and reseal the foil pouch; use within one month.

What is the detection range of the assay?

The assay can detect SOD2 concentrations from 0.78 ng/mL up to 50 ng/mL.