Sauermann Si-TI3 non-contact infrared thermometer user manual covering specifications, operation, emissivity table, maintenance, and mobile app.
Sauermann Si-TI3 is a non-contact infrared thermometer that measures infrared surface temperature and ambient temperature. It has an infrared range of -40 to +500°C, adjustable emissivity from 0.10 to 1.0, a 12:1 distance-to-spot ratio, and a backlit LCD.
The manual covers safety, battery installation, measurement, unit selection, emissivity setting, max/min/difference/average modes, alarms, laser sighting, storage, dimensions, accessories, maintenance, precautions, and the Si-HVACR Measurement MobileApp. This manual is written in Chinese.
| Infrared temperature range | -40 to +500°C |
|---|---|
| Infrared temperature accuracy | ±(2% of reading +2°C) |
| Infrared temperature resolution | 0.1°C |
| Ambient temperature range | 0 to 50°C |
| Emissivity | Adjustable 0.10-1.0, default 0.95 |
| Response time | T90 1 s |
| Distance-to-spot ratio | 12:1 |
| Laser sighting | Single point |
| Display | 3-line backlit black-and-white LCD |
| Housing material | ABS-PC plastic backplate |
| Power supply | 2 AAA 1.5 V alkaline batteries |
| Battery life | 14 hours (backlight off) |
| Weight | 145 g including battery |
| Auto power off | 15 s |
| Operating conditions | 0-50°C, non-condensing, altitude 0-2000 m |
| Storage temperature | -10 to +60°C |
| Wireless communication | RF 2402-2480 MHz, range up to 15 m |
| Minimum app requirements | Android 5.0, iOS 12.4, Bluetooth 4.0 |
| Standards | RoHS II, WEEE, RED |
| EMC/RFI | Max 10°C error in 200-600 MHz, 3 V/m field |
| Laser | Class 2, >1 mW, 635-660 nm |
| Dimensions | 112.6 x 148.2 x 41.3 mm |
The default emissivity is 0.95, and it is adjustable from 0.10 to 1.0.
Battery life is 14 hours with the backlight off.
The Si-HVACR Measurement MobileApp can view and record measurements in real time.
Wait 15 seconds for automatic shutoff, remove the battery, wait at least 1 minute, reinsert the battery, and press any key to turn it on.
Er2 appears when the instrument is exposed to a rapidly changing environment, and Er3 appears when ambient temperature is outside 0°C to +50°C.
Wipe it carefully with a soft cloth or cotton swab moistened with water or medical alcohol, let it dry completely, and do not immerse any part of the thermometer in water.