This manual introduces the use of Sauermann temperature and humidity measurement instrument Si-HH3, precautions and safety use specifications.
This is the Chinese operating manual for the Sauermann Si-HH3 thermo-hygrometer, a handheld instrument that measures relative humidity and temperature and also calculates dew point, absolute humidity, enthalpy, mixing ratio and wet-bulb temperature. The document describes the humidity probe, the backlit two-line LCD, the on/off and backlight key and the HOLD key, then lists operating environment and safety instructions, the FCC Part 15 statement and the full technical specifications.
The manual continues with a dimension drawing and step-by-step operating instructions: inserting the three AAA batteries, switching the instrument on, selecting °C or °F, turning the wireless connection and backlight on or off, changing the auto power-off setting, and cycling through hold, minimum, maximum and real-time values. A maintenance section covers battery replacement and cleaning, followed by a note on the 2-year warranty and a short description of the Si-HVACR Measurement MobileApp. This manual is written in Chinese.
| Humidity sensor | Capacitive sensor |
|---|---|
| Display | Integrated 2-line backlit monochrome LCD |
| Housing / protection | ABS-PC plastic back plate, IP20 |
| Power supply | 3 x 1.5 V AAA alkaline batteries |
| Power consumption | 135 mW |
| Battery life | 250 hours (backlight and wireless off) |
| Wireless connection | 2402 MHz - 2480 MHz at 0 dBm, range up to 15 m |
| Operating conditions | -10 to +60 °C, non-condensing, altitude 0 - 2000 m |
| Storage temperature | -20 to +70 °C |
| Auto power off | 10 minutes (can be disabled) |
| Weight | 200 g including batteries |
| Relative humidity measurement | 0 - 100 %RH, accuracy ±2 %RH, resolution 0.1 %RH |
| Temperature measurement | -20 to +60 °C, accuracy ±0.5 °C, resolution 0.1 °C |
| Dimensions | 228 mm x 70 mm x 33 mm, probe Ø13 mm |
With the instrument switched on, press the on/off key and the HOLD key at the same time to cycle to the required unit. Selecting °F also displays °FTd and °FTw, and selecting °C also displays °CTd and °CTw.
The stated battery life is 250 hours with the backlight and wireless communication not in use, measured at 20 °C with alkaline batteries.
With the device switched off, unscrew the battery cover on the back, open it, take out the old batteries and fit three new 1.5 V AAA alkaline batteries, observing polarity, then refit the cover from top to bottom and tighten the screw.
Wipe the housing with a damp cloth and a little soap solution; do not use corrosive cleaners or solvents.