Guide for verifying calibration of Sage Insertion Thermal Mass Flow Meters using in-situ 'No Flow' (0 SCFM) conditions.
This manual explains how to verify the calibration of Sage Insertion Thermal Mass Flow Meters by creating a 'No Flow' (0 SCFM) condition. It details methods using SVA05 or SVA07 Isolation Valve Assemblies, bypass pipes, or removal for zeroing. The procedure confirms sensor cleanliness and calibration integrity, complying with U.S. EPA 40 CFR Part 98 and landfill/livestock protocols.
The Sage Zeroing Chamber allows capturing a 'No Flow' condition under actual process conditions. Comparing observed milliwatts (mW) to factory data verifies reproducibility. The digital sensor technology enables in-situ verification, eliminating annual calibrations for many users.
You can create a no-flow condition by temporarily shutting down the flow, valving it off, installing a bypass pipe, or conducting the check during a planned maintenance shutdown.
The verification procedure requires creating a condition of 0 SCFM to check the meter's reproducibility.
You should compare the observed milliwatts to the factory's original data point taken at a no-flow condition.