The Roche cobas c 111 is a compact clinical chemistry analyzer for labs processing up to 50 samples per day, offering photometric and ISE testing with…
The cobas c 111 analyzer is the smallest member of Roche's cobas platform family, designed for laboratories processing up to 50 samples per day. It provides a compact solution for clinical chemistry, electrolyte, and homogeneous immunoassay testing, suitable for small labs, private labs, satellite labs, and as a backup or dedicated STAT analyzer.
Key features include a dynamic transfer head that performs four functions in one, optional ISE unit for indirect measurement of chloride, potassium, and sodium, continuous sample loading with STAT priority, exchangeable reagent disks for flexible panel testing, low water consumption up to 3 liters per day, and an intuitive touchscreen interface with guided start-up and maintenance. The system delivers comparable results across the cobas platform using common reagent chemistry, calibrators, and controls.
| Test throughput | 60-85 photometric tests/hr, 180 ISE tests/hr, 60-100 mixed tests/hr |
|---|---|
| Sample types | Serum, plasma, urine, whole blood (HbA1c) |
| Sample input | Continuous loading into 8 sample positions with STAT priority |
| Time to first result | 5-10 min photometric, 2 min ISE |
| Sample volume | Primary tubes 500 µL, sample cup 75 µL, micro cup 50 µL |
| Photometer | 12 wavelengths, 20 W halogen lamp, mono- and bichromatic |
| Measurement principles | Absorbance photometry and indirect ISE (dilution 1:6) |
| Reagents | ISE: Na+, K+, Cl-; photometric: 27 onboard positions, up to 8 exchangeable discs |
| Control unit | 5.7" color touch-screen LCD (1/4 VGA) |
| System interfaces | 2x RS-232 (ASTM), 2x USB 1.1/2.0 |
| Electrical requirements | 100-125 V / 200-240 V AC, 50/60 Hz, 250 VA (320 VA with ISE) |
| Physical dimensions (W×D×H) | Without ISE: 590×550×480 mm; with ISE: 720×550×480 mm |
| Weight | 32.3 kg (35 kg with ISE) |
| Water consumption | Up to 3 L/day, NCCLS Type II |
| Regulatory compliance | GS, CE, UL, C-UL |
The analyzer delivers 60-85 photometric tests per hour, 180 ISE tests per hour, and 60-100 mixed tests per hour.
It accepts serum, plasma, urine, and whole blood for HbA1c testing.
It operates on 100-125 V or 200-240 V AC at 50/60 Hz, with power consumption of 250 VA (320 VA with ISE).
Without ISE: 590 mm wide, 550 mm deep, 480 mm high; with ISE: 720 mm wide, 550 mm deep, 480 mm high.