Roche cobas c 111 Clinical Chemistry Analyzer Product Brochure

The Roche cobas c 111 is a compact clinical chemistry analyzer for labs processing up to 50 samples per day, offering photometric and ISE testing with…


Roche cobas c 111 Clinical Chemistry Analyzer Product Brochure - cover page
Brand
Roche
Category
Analyzer
Document type
Product Leaflet
Language
English
Pages
14
File format
PDF
File size
979 KB
Published
07 November, 2026
Updated
07 November, 2026
MD5 checksum
1EDFFDFF7CEC59BF5056FBED34E71387

About this manual

The cobas c 111 analyzer is the smallest member of Roche's cobas platform family, designed for laboratories processing up to 50 samples per day. It provides a compact solution for clinical chemistry, electrolyte, and homogeneous immunoassay testing, suitable for small labs, private labs, satellite labs, and as a backup or dedicated STAT analyzer.

Key features include a dynamic transfer head that performs four functions in one, optional ISE unit for indirect measurement of chloride, potassium, and sodium, continuous sample loading with STAT priority, exchangeable reagent disks for flexible panel testing, low water consumption up to 3 liters per day, and an intuitive touchscreen interface with guided start-up and maintenance. The system delivers comparable results across the cobas platform using common reagent chemistry, calibrators, and controls.

What's inside

  • Designed to fit in many ways
  • High level performance on a small footprint
  • Integrated service and support
  • World-class performance
  • Good fit for labs processing up to 50 samples per day
  • Designed to complete the network offering
  • Specifications
  • Serum Work Area test menu

Specifications

Test throughput60-85 photometric tests/hr, 180 ISE tests/hr, 60-100 mixed tests/hr
Sample typesSerum, plasma, urine, whole blood (HbA1c)
Sample inputContinuous loading into 8 sample positions with STAT priority
Time to first result5-10 min photometric, 2 min ISE
Sample volumePrimary tubes 500 µL, sample cup 75 µL, micro cup 50 µL
Photometer12 wavelengths, 20 W halogen lamp, mono- and bichromatic
Measurement principlesAbsorbance photometry and indirect ISE (dilution 1:6)
ReagentsISE: Na+, K+, Cl-; photometric: 27 onboard positions, up to 8 exchangeable discs
Control unit5.7" color touch-screen LCD (1/4 VGA)
System interfaces2x RS-232 (ASTM), 2x USB 1.1/2.0
Electrical requirements100-125 V / 200-240 V AC, 50/60 Hz, 250 VA (320 VA with ISE)
Physical dimensions (W×D×H)Without ISE: 590×550×480 mm; with ISE: 720×550×480 mm
Weight32.3 kg (35 kg with ISE)
Water consumptionUp to 3 L/day, NCCLS Type II
Regulatory complianceGS, CE, UL, C-UL

Frequently asked questions

What is the test throughput of the cobas c 111 analyzer?

The analyzer delivers 60-85 photometric tests per hour, 180 ISE tests per hour, and 60-100 mixed tests per hour.

What sample types can the cobas c 111 analyzer process?

It accepts serum, plasma, urine, and whole blood for HbA1c testing.

What are the electrical requirements for the cobas c 111 analyzer?

It operates on 100-125 V or 200-240 V AC at 50/60 Hz, with power consumption of 250 VA (320 VA with ISE).

What are the physical dimensions of the cobas c 111 analyzer?

Without ISE: 590 mm wide, 550 mm deep, 480 mm high; with ISE: 720 mm wide, 550 mm deep, 480 mm high.