This manual covers the operation, safety precautions, maintenance, and specifications of the RIKENKEIKI RI-557 portable infrared gas detector for measuring…
The RIKENKEIKI RI-557 is a portable infrared gas detector designed to measure the concentration of specific gases such as carbon monoxide, carbon dioxide, and combustible gases. It utilizes an internal pump to draw atmospheric samples for detection and provides both concentration signal output and alarm contact outputs based on preset levels. The device is intended for monitoring environments like research facilities and plastic greenhouses where these gases are used or generated.
This manual details the safe operation of the instrument, including startup procedures, flow adjustment, and zero/span calibration. It also outlines safety warnings, such as the prohibition of using the device in explosive atmospheres or for controlling external equipment, and provides instructions for maintenance, troubleshooting, and handling abnormal conditions. This manual is written in Japanese.
| Power Supply | AC100V 50/60Hz or AC220V 50/60Hz |
|---|---|
| Terminal Rated Voltage | AC250V |
| Terminal Rated Current | 15A |
| Terminal Screw Size | M4 |
| Maximum Cable Size | 1.25mm² |
| Signal Output Range | 4~22mA (or 0-1V) |
| Zero/Span Mode Output | 2.5mA (or 0V) |
| Fault Alarm Output | 0.5mA (or 0V) |
No, the instrument is a safety device and should not be used to control other equipment.
It is recommended to perform zeroing approximately once a month.
Regular inspections and gas calibration are recommended every 6 months according to regulations.
No, the alarm contacts are for signaling external alarms or indicators, not for controlling loads like valves.