Simplified Chinese user guide for the ResMed H4i heated humidifier used with compatible CPAP, bilevel and non-invasive ventilation devices, covering setup…
This is the Simplified Chinese user guide for the ResMed H4i heated humidifier, which is designed to humidify air delivered to the airway during positive airway pressure therapy. It is intended for use only with compatible ResMed CPAP, bilevel and non-invasive ventilation devices and only as prescribed.
The guide explains how to assemble and set up the H4i with a flow generator, fill the water chamber, start and stop therapy including warm-up with VPAP, S8 series and Stellar devices, and adjust the humidity control dial. It also covers daily and weekly cleaning of the water chamber, lid and seal, disassembly and reassembly of the cleanable water tub, troubleshooting, technical specifications, warnings and the limited warranty. This manual is written in Chinese.
| Heater plate maximum temperature | 85°C |
|---|---|
| Cut-out temperature | 93°C |
| Maximum air temperature | 41°C |
| Dimensions (H x W x D) | 130 x 170 x 212 mm (base and water tub) |
| Weight | 786 g (base and tub, without water) |
| Water tub capacity | 390 mL to maximum water level |
| Power supply | 100–240 V, 50/60 Hz, 0.7–1.6 A; 110 V, 400 Hz, 0.8 A; peak power <400 W, average power <115 W |
| Operating temperature | +5°C to +35°C |
| Operating humidity | 10–95%, non-condensing |
| Storage and transport temperature | -20°C to +60°C |
| IEC 60601-1 classification | Class II (double insulation), BF-type applied part |
Condensation can occur when the room temperature is low. Turn the control dial down to a lower setting, adjust the humidity setting as seasons change, and using a ResMed tubing wrap can help reduce condensation.
No, the H4i must not be used when the treatment pressure exceeds 35 cmH2O.
The disassembled cleanable water tub, spill guard, base plate and base plate seal can be washed on the top shelf of a dishwasher using a gentle or glassware cycle; this does not apply to the standard H4i water tub.
The H4i should be inspected by an authorised ResMed service centre five years from the date of manufacture.