Operating and maintenance guidelines for Precision vaporizers used in inhalation anesthesia, covering preparation, installation, and usage.
This document provides operating instructions and maintenance guidelines for Precision vaporizers used in inhalation anesthesia. It includes warnings, cautions, descriptions, specifications, performance details, and installation and operation instructions. The manual emphasizes that the vaporizer is designed for 'out of circuit' use in continuous flow techniques and is both flow rate and temperature compensated.
| Calibration Temperature | 21°C (70°F) |
|---|---|
| Flow Compensation Range | 300 ml to 10 liters per minute |
| Resistance to Flow | 40 cm/H2O (30 mm/Hg) at 10 liters per minute |
No, do not fill the vaporizer with any anesthetic agent other than the one specified on the front label.
The vaporizer must be in a vertical upright position to indicate liquid agent levels accurately.
An annual calibration recertification is highly recommended by the manufacturer.