Physio-Control LIFEPAK CR2 Automated External Defibrillator User Manual

Physio-Control LIFEPAK CR2 AED manual covering intended use, contraindications, warnings, battery maintenance, cleaning and product specifications.


Physio-Control LIFEPAK CR2 Automated External Defibrillator User Manual - cover page
Brand
Physio Control
Category
Professional medical equipment
Document type
User Manual
Language
Chinese
Pages
21
File format
PDF
File size
1.1 MB
Published
10 November, 2026
Updated
10 November, 2026
MD5 checksum
1F3E0AF8CFE79C8EDDB339190E7B4074

About this manual

This document describes the Physio-Control LIFEPAK CR2 automated external defibrillator (AED), a device used to respond to sudden cardiac arrest. After electrode pads are placed on the patient's chest, the device analyzes the heart rhythm and, if a shockable rhythm is detected, either delivers an electric pulse automatically (fully automatic mode) or prompts the rescuer to press the shock button (semi-automatic mode). Voice prompts guide the rescuer through the event in both modes. The device can also connect over Wi-Fi, mobile network or USB to the LIFELINKcentral AED Program Manager or a LIFENET System account (not included with this registration); the micro-USB port is for authorized users only and cannot supply power.

The document states that the device is intended for patients 1 year of age and older in cardiopulmonary arrest and may only be used by personnel trained in its operation, together with QUIK-STEP therapy/ECG/defibrillation electrodes and the LIFEPAK CR2 lithium battery. It also sets out contraindications, general hazards and warnings, battery maintenance and device cleaning instructions, and product specifications. A new battery gives about 800 minutes of on time or 166 200-joule discharges, and a battery installed in an unused device remains effective for 4 years. This manual is written in Chinese.

What's inside

  • Product Description
  • Intended Use
  • Contraindications
  • Warnings and Cautions
  • Using the Defibrillator
  • Maintenance and Care
  • Battery Maintenance
  • Cleaning the Device
  • Specifications

Specifications

WaveformBiphasic truncated exponential with voltage and shock duration compensation for patient impedance
Patient impedance range10 to 300 ohms
Output energy, adult mode150, 200, 300 or 360 joules
Output energy, pediatric mode35, 50, 75 or 90 joules
Maximum delivered energyUp to 360 joules
Battery capacityAbout 800 minutes of on time or 166 200-joule discharges
Battery life when installed and unused4 years
Battery typeLIFEPAK CR2 non-rechargeable lithium manganese dioxide battery
Applicable patient age1 year and older
StandardsInternally powered equipment per IEC 60601-1

Frequently asked questions

What is the LIFEPAK CR2 used for?

It is an automated external defibrillator (AED) used to respond to sudden cardiac arrest; after electrodes are attached, it analyzes the patient's rhythm and delivers or prompts a shock when a shockable rhythm is detected.

What is the difference between fully automatic and semi-automatic mode?

In fully automatic mode the device delivers the shock without rescuer action once a shockable rhythm is detected; in semi-automatic mode the rescuer must press the shock button.

What accessories does the LIFEPAK CR2 use?

It is used with QUIK-STEP therapy/ECG/defibrillation electrodes and the LIFEPAK CR2 lithium battery.

Who is allowed to use the device?

It may only be used by personnel who have been trained in its operation, such as those trained in BLS/AED or ALS.

How should the device be cleaned?

Clean the defibrillator after use with a damp sponge or cloth using only non-abrasive soap and water, quaternary ammonium compounds, wiping (isopropyl) alcohol, or peracetic acid solution.