Physio-Control LIFEPAK CR2 AED manual covering intended use, contraindications, warnings, battery maintenance, cleaning and product specifications.
This document describes the Physio-Control LIFEPAK CR2 automated external defibrillator (AED), a device used to respond to sudden cardiac arrest. After electrode pads are placed on the patient's chest, the device analyzes the heart rhythm and, if a shockable rhythm is detected, either delivers an electric pulse automatically (fully automatic mode) or prompts the rescuer to press the shock button (semi-automatic mode). Voice prompts guide the rescuer through the event in both modes. The device can also connect over Wi-Fi, mobile network or USB to the LIFELINKcentral AED Program Manager or a LIFENET System account (not included with this registration); the micro-USB port is for authorized users only and cannot supply power.
The document states that the device is intended for patients 1 year of age and older in cardiopulmonary arrest and may only be used by personnel trained in its operation, together with QUIK-STEP therapy/ECG/defibrillation electrodes and the LIFEPAK CR2 lithium battery. It also sets out contraindications, general hazards and warnings, battery maintenance and device cleaning instructions, and product specifications. A new battery gives about 800 minutes of on time or 166 200-joule discharges, and a battery installed in an unused device remains effective for 4 years. This manual is written in Chinese.
| Waveform | Biphasic truncated exponential with voltage and shock duration compensation for patient impedance |
|---|---|
| Patient impedance range | 10 to 300 ohms |
| Output energy, adult mode | 150, 200, 300 or 360 joules |
| Output energy, pediatric mode | 35, 50, 75 or 90 joules |
| Maximum delivered energy | Up to 360 joules |
| Battery capacity | About 800 minutes of on time or 166 200-joule discharges |
| Battery life when installed and unused | 4 years |
| Battery type | LIFEPAK CR2 non-rechargeable lithium manganese dioxide battery |
| Applicable patient age | 1 year and older |
| Standards | Internally powered equipment per IEC 60601-1 |
It is an automated external defibrillator (AED) used to respond to sudden cardiac arrest; after electrodes are attached, it analyzes the patient's rhythm and delivers or prompts a shock when a shockable rhythm is detected.
In fully automatic mode the device delivers the shock without rescuer action once a shockable rhythm is detected; in semi-automatic mode the rescuer must press the shock button.
It is used with QUIK-STEP therapy/ECG/defibrillation electrodes and the LIFEPAK CR2 lithium battery.
It may only be used by personnel who have been trained in its operation, such as those trained in BLS/AED or ALS.
Clean the defibrillator after use with a damp sponge or cloth using only non-abrasive soap and water, quaternary ammonium compounds, wiping (isopropyl) alcohol, or peracetic acid solution.