User manual for the Philips CM 100 electron microscope, covering startup, alignment, sample preparation, insertion, PC control, focusing, and image saving.
This user manual covers the Philips CM 100 electron microscope. It gives step-by-step instructions for starting the microscope, performing beam alignment, preparing and inserting samples, and shutting the system down.
It also describes starting the PC and the Gatan Digital Micrograph camera software, adjusting the Z-axis, focusing images, and saving images as TIFF or DM3 files.
| Magnification | 2500x |
|---|---|
| Magnification | 7400x and 9700x |
Push PANEL DIM to ON, turn on DATA DIM, push HIGH TENSION ON, rotate FILAMENT clockwise 4-5 times, and wait 3 minutes.
Keep the holder horizontal, insert it forward to the first stop until a red light appears on the goniometer, activate ST CRYO HOLDER, confirm READY, wait for the red light to turn off, then turn counter-clockwise until it is sucked in.
Images can be saved as TIFF or DM3 files in your folder under ZMB.
Log off the computer and shut down the microscope by carrying out the starting process in reverse order.