This document provides specifications and features for the Panametrics 5072PR, 5073PR, and 5077PR ultrasonic pulser-receivers used in flaw detection and…
This document details the specifications and features of three manually controlled ultrasonic pulser-receiver models: the 5072PR, 5073PR, and 5077PR. These instruments are designed as building blocks for ultrasonic test systems, offering optimization for both conventional and high-frequency applications. They are used with transducers and oscilloscopes for flaw detection, thickness gaging, and material characterization.
| Bandwidth (–3 dB, RL = 50 Ω) 5072PR | 1 kHz to 35 MHz |
|---|---|
| Bandwidth (–3 dB, RL = 50 Ω) 5073PR | 1 kHz to 75 MHz |
| Bandwidth (–3 dB, RL = 50 Ω) 5077PR | 1 kHz to 35 MHz |
| Pulse type (main bang) 5072PR | Negative impulse |
| Pulse type (main bang) 5073PR | Negative |
| Pulse type (main bang) 5077PR | Negative square wave |
| Rise time (10 % to 90 %) 5072PR | Typically 5 ns, 10 ns max. |
| Rise time (10 % to 90 %) 5073PR | Typically < 2 ns |
| Rise time (10 % to 90 %) 5077PR | Typically < 10 ns, 20 ns max. |
| Available pulse voltage (no load) 5077PR | –400 V, –300 V, –200 V, –100 V selectable |
| Operating temperature range | 0 °C to 50 °C (32 °F to 122 °F) |
| Dimensions (W x H x D) | 178 mm x 89 mm x 232 mm (7 in. x 3.5 in. x 9.1 in.) |
| Weight | 2.3 kg (5 lb) |
The 5073PR has a bandwidth of 1 kHz to 75 MHz.
The 5077PR uses a negative square wave pulse type.
The maximum rise time for the 5072PR is 10 ns.