Manual for the Onset S-THB-M00x temperature/RH smart sensor, including specifications, mounting, connection, RH sensor replacement, and maintenance.
The S-THB-M00x temperature/RH smart sensor is designed to work with smart sensor-compatible HOBO data loggers and stations. Sensor parameters are stored inside the smart sensor and are communicated automatically to the logger.
The manual provides temperature and RH specifications, mounting and connection guidance, steps for replacing the RH sensor, and maintenance recommendations for the sensor.
| Temperature accuracy | ±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F) |
|---|---|
| RH accuracy | ±2.5% from 10% to 90% RH typical to a maximum of ±3.5% including hysteresis at 25°C (77°F); below 10% and above 90% ±5% typical |
| Temperature resolution | 0.02°C at 25°C (0.04°F at 77°F) |
| Bits per sample | 12 (temperature), 10 (RH) |
| Drift | <0.1°C (0.18°F) per year; <1% per year typical |
| Response time | 5 minutes in air moving 1 m/sec |
| Operating temperature range | -40°C to 75°C (-40°F to 167°F) |
| Environmental rating | Weatherproof: 0 to 100% RH intermittent condensing environments |
| Housing | PVC cable jacket with ASA styrene polymer RH sensor cap; modified hydrophobic polyethersulfone membrane |
| Sensor dimensions | 10 x 35 mm (0.39 x 1.39 in) |
| Weight | S-THB-M002: 110 g (3.88 oz); S-THB-M008: 180 g (6.35 oz) |
| Number of data channels | 2 |
| Measurement averaging option | No |
| Cable lengths available | 2.5 m (8.2 ft); 8 m (26.2 ft) |
| Smart sensor network cable length | 0.5 m (1.6 ft); 6 m (19.6 ft) |
The sensor is available with 2.5 m (8.2 ft) or 8 m (26.2 ft) cables.
It provides 2 data channels.
Remove the cap and membrane, remove the old sensor board, install the new HUM-RHPCB-2 sensor in the same orientation, then fit the new cap and membrane.
Periodically inspect the RH sensor; rinse the protective cap with cool fresh water or the sensor with distilled water, and avoid hot water, organic solvents, or detergents.