User manual for the Obelab NIRSIT NS1-H20AM optical topography system, a non-invasive device that continuously monitors cerebral oxygen saturation (rSO2).
This is the Traditional Chinese instructions-for-use booklet for the Obelab NIRSIT (NS1-H20AM), an optical topography system that measures blood oxygen saturation through the skin with optical sensor probes. Using near-infrared spectroscopy (NIRS), it continuously and non-invasively monitors cerebral oxygen saturation (rSO2) in adults and in children aged 8 and above; the monitor has two channels that work independently to measure and display oxygen saturation distribution, and the values are shown on the monitor. The document carries the Taiwanese registration number 衛部醫器輸字第 034346號.
The booklet covers the product description and intended use, contraindications, warnings and safety precautions, specifications, package contents, cleaning and maintenance, and labeled part names with illustrations. It states that the device is for medical use only, must be operated by medical personnel or fully trained persons, and must be used together with the dedicated software NIRSIT-EYE; the supplied tablet is restricted to this product. Manufacturer information for Obelab, Inc. (Seoul, Korea) and the local distributor 春杏醫療儀器有限公司 is given at the end. This manual is written in Chinese.
| Model | NIRSIT (NS1-H20AM) |
|---|---|
| Medical device license | 衛部醫器輸字第 034346號 |
| rSO2 measurement range / accuracy | 15%~95%, ±4% |
| Dimensions (W x D x H) | 216 x 191 x 77.8 mm |
| Weight (including battery) | 500 g |
| Emitter type | Dual-wavelength VCSEL laser light, 24 units |
| Laser wavelength | 780 nm, 850 nm |
| Total channels | 204 (maximum) |
| System scan rate | 8.138 Hz |
| Input / maximum current | 5 V, 2.1 A (USB connector) |
| Battery | Lithium-ion polymer, 3.7 V, 3000 mAh, up to 8 hours use |
| Application site and contact time | Forehead, up to 60 minutes continuous contact |
| Operating conditions | 15 °C to 35 °C, 20% to 80% humidity |
| Storage conditions | -10 °C to 40 °C, 20% to 80% humidity |
| Communication | Wireless WLAN (2.4 GHz 802.11b/g/n); wired serial communication |
It uses near-infrared spectroscopy to measure blood oxygen saturation under the sensor and continuously monitors cerebral oxygen saturation (rSO2) non-invasively in adults and children aged 8 and above.
The device is for medical use only and must be used by medical personnel or personnel who have been fully trained, together with the dedicated software.
Continuous contact at the forehead is limited to 60 minutes or less.
Wipe the silicone cover with a soft cloth moistened with alcohol; because it frequently contacts the subject, special attention to hygiene is required.