This operation manual describes the Nuvair Pro CO2 Carbon Dioxide Analyzer, a battery-powered NDIR instrument that measures CO2 in gases in the range of 0 to 2000 ppm. It is designed to verify CO2 concentration in stored gas cylinders and to monitor continuous flow of sample gas from a compressor, and it is used with a flow adapter cap and flexible tubing that deliver sample gas to the sensor.
The manual explains the controls, display, alarm, sensor, battery and flow adapter cap, then walks through testing, sample flow method, programming procedures, alarm setting, full scale value setting, calibration point setting and zero calibration. It also covers threshold alarms, powering off, factory reset, maintenance such as analyzer care, battery and sensor replacement, spares and accessories, troubleshooting, relay output schematics, and an appendix with analyzer specifications and warranty information.
| Measuring Range | 0 to 2000 ppm |
|---|---|
| Resolution | 0 – 2500 (50 ppm) |
| Alarm Set Point | Customer set ppm of CO2 |
| Display Accuracy | +/- 5% |
| Sensor Type | NDIR non-dispersive infrared |
| Expected Sensor Life, Room Air | 5 years |
| Power | Lithium battery or optional 110/230 V electric |
| Battery Life | 6 hours |
| Response Time | Less than 50 seconds to 90% of final value |
| Operating Temperature | 41 to 104 °F (5 to 40 °C) |
| Storage Temperature | 14 to 140 °F (-10 to 60 °C) |
| Operating Pressure | Not to exceed 1 atmosphere absolute (0 P.S.I.) |
| Humidity | 15-90% continuous, 0-99% intermittent |
Hold the On/Off button until the display powers up to switch it on, and hold it until the display goes blank to turn it off.
Gas flow should be regulated to 0.5-3 L/min with no pressure on the sensor; for sample flow through a Nuvair 9517.6 flow reducer the incoming pressure should be 75-100 PSI.
The audible alarm activates when the sensor reaches 199 ppm CO2 or a user programmed level, and it will not clear until the CO2 concentration drops below the user programmed level.
Analyzer testing must be verified on a weekly basis, or when moisture and/or temperature swings take place.