NevadaNano MPS Flammable Gas Sensor 4.0 Series 4 (S4) User Manual

User manual for the NevadaNano MPS Flammable Gas Sensor 4.0 Series 4 (S4), covering operation, UART communication, analog output, integration, testing, and…


NevadaNano MPS Flammable Gas Sensor 4.0 Series 4 (S4) User Manual - cover page
Brand
NEVADANANO
Category
sensor
Document type
User Manual
Language
English
Pages
35
File format
PDF
File size
864 KB
Published
17 November, 2026
Updated
17 November, 2026
MD5 checksum
2A41A847DE02FAD75CAAACBF817F2C07

About this manual

The MPS Flammable Gas Sensor 4.0 is a smart sensor with built-in environmental compensation that detects and accurately quantifies over a dozen gases and gas mixtures. It is intrinsically safe, poison-resistant, and calibrated for all gases by performing a single calibration with methane; readings are output on a standard digital bus or industry-standard analog output.

This manual describes UART communication, packet protocol, status codes, analog output configurations, electromechanical integration, test setups, environmental faults, hazardous-location considerations, sensor data, and warranty information.

What's inside

  • How the MPS Flammable Gas Sensor 4.0 Works
  • Serial (UART) Communication
  • Sensor Analog Output Standard Mode
  • Custom Analog Out Configurations
  • Analog Out Load Resistance
  • Electromechanical Integration
  • Testing the MPS Flammable Gas Sensor 4.0
  • Environmental Faults and Conditions Affecting Accuracy
  • Hazardous Locations and System Integration
  • Warranty information

Specifications

Serial baud rate38400
Serial data size8 bits
Serial parityNone
Serial stop bit1 bit
Supply voltage3.3-5.0 VDC ±5%

Frequently asked questions

What gases can the MPS Flammable Gas Sensor 4.0 detect?

It detects and accurately quantifies over a dozen gases as well as gas mixtures.

What is the serial baud rate for communicating with the sensor?

The serial connection uses a baud rate of 38400.

What status is returned when transmitted data fails the checksum?

Status 0x01 CRC_FAILED is returned, and the checksum calculation should be verified.

What supply voltage is required for the sensor?

The sensor requires a supply of 3.3-5.0 VDC ±5%.

Download PDF (864 KB)