The Natus Quantum Amplifier EEG recorder labeling covers intended use, contraindications, safety warnings, specifications, and component names.
The Natus Quantum Amplifier is an EEG recorder system intended to capture, display, record, and store electroencephalographic and physiological sleep data. It connects to software and a computer to process EEG and polysomnographic recordings; the document states that the product does not include an isolation transformer.
The document lists contraindications and safety precautions, including restrictions for use with flammable anesthetics, MRI or CT systems, and certain electrical and patient-environment conditions. It also provides analog and digital specifications and lists the Base Unit, Main Breakout, Breakout 2, and breakout cables. This manual is written in Chinese.
| Common-mode rejection ratio (CMRR) | ≥110 dB @ 60 Hz |
|---|---|
| Common-mode input impedance | >1 GΩ |
| Differential input range | 20 mVpp |
| Input bias current | <1 nA |
| Sampling frequency | 256, 512, 1024, 2048, 4096, 8192, 16384 Hz |
| Bandwidth | 0.01 Hz to 4 kHz |
| Digital/analog sampling resolution | 24 bits |
| Power and input voltage | 70 VA; 100–230 V, 50/60 Hz |
| Base Unit dimensions | 29.0 × 26.5 × 5.0 cm (L × W × H) |
| Breakout Box weight | 650 g |
| Operating temperature | 10–40 °C |
| Operating relative humidity | 30–75% |
It captures, displays, records, and stores EEG and physiological sleep data, and works with software and a computer to process EEG and PSG signals.
No. The contraindications state that the product must not be used on a fetus.
No. The document states that the product does not include the power supply component (Isolation Transformer).
Available sampling frequencies are 256, 512, 1024, 2048, 4096, 8192, and 16384 Hz.