Natus Quantum Amplifier EEG Recorder Product Leaflet

The Natus Quantum Amplifier EEG recorder labeling covers intended use, contraindications, safety warnings, specifications, and component names.


Natus Quantum Amplifier EEG Recorder Product Leaflet - cover page
Brand
natus
Category
Professional medical equipment
Document type
Product Leaflet
Language
Chinese
Pages
8
File format
PDF
File size
355 KB
Published
16 November, 2026
Updated
16 November, 2026
MD5 checksum
5459CBAFB338AB21B1C38700F7B3F235

About this manual

The Natus Quantum Amplifier is an EEG recorder system intended to capture, display, record, and store electroencephalographic and physiological sleep data. It connects to software and a computer to process EEG and polysomnographic recordings; the document states that the product does not include an isolation transformer.

The document lists contraindications and safety precautions, including restrictions for use with flammable anesthetics, MRI or CT systems, and certain electrical and patient-environment conditions. It also provides analog and digital specifications and lists the Base Unit, Main Breakout, Breakout 2, and breakout cables. This manual is written in Chinese.

What's inside

  • Product Description
  • Intended Use
  • Contraindications
  • Warnings and Precautions
  • Specifications
  • Analog Specifications – AC Channels
  • Analog Specifications – DC Channels
  • Digital Specifications
  • Operating Environment Conditions
  • Product Components and Names

Specifications

Common-mode rejection ratio (CMRR)≥110 dB @ 60 Hz
Common-mode input impedance>1 GΩ
Differential input range20 mVpp
Input bias current<1 nA
Sampling frequency256, 512, 1024, 2048, 4096, 8192, 16384 Hz
Bandwidth0.01 Hz to 4 kHz
Digital/analog sampling resolution24 bits
Power and input voltage70 VA; 100–230 V, 50/60 Hz
Base Unit dimensions29.0 × 26.5 × 5.0 cm (L × W × H)
Breakout Box weight650 g
Operating temperature10–40 °C
Operating relative humidity30–75%

Frequently asked questions

What does the Natus Quantum Amplifier do?

It captures, displays, records, and stores EEG and physiological sleep data, and works with software and a computer to process EEG and PSG signals.

Can the Natus Quantum Amplifier be used on a fetus?

No. The contraindications state that the product must not be used on a fetus.

Does the system include an isolation transformer?

No. The document states that the product does not include the power supply component (Isolation Transformer).

What sampling frequencies are available?

Available sampling frequencies are 256, 512, 1024, 2048, 4096, 8192, and 16384 Hz.