Calibration operation manual for the MinebeaMitsumi MMR920, covering calibration data updates, P result max/min settings, slope calculation, calibration…
This manual describes the calibration operation for the MinebeaMitsumi MMR920. It covers updating calibration data, setting P result maximum and minimum values, calculating slope and offset, enabling calibration, and the calibration log format.
The described procedure includes applying maximum or minimum pressure, starting measurement, reading the pressure measurement value when stable, and optional manual entry. It also gives the equations for calibrated results, slope, and offset.
Apply maximum or minimum pressure, start measurement, and read the pressure measurement value when it is stable; manual entry is also possible.
The calibrated result is calculated as P_Result_Calib = P_Result × Slope + Offset.
Slope = (P_Max - P_Min) / (P_Result_Max - P_Result_Min), and Offset = P_Min - P_Result_Min × Slope.
The log format shows calibrated data and raw data for calibration on and calibration off states.