User manual for the MEIRUIKE RK9914 Series AC/DC withstand voltage tester, covering safety, installation, operation, parameter settings, serial commands…
The MEIRUIKE RK9914 Series is a program-controlled AC/DC withstand voltage tester. The user manual covers installation and use, operating precautions, front and rear panel descriptions, operation steps, parameter settings, system settings, test function principles, and HANDLER/SIGNAL interface circuits.
It also documents the serial command set, including SCPI and subsystem commands, and provides instrument specifications and a packing list. Safety warnings cover grounding, electric shock, high-voltage testing, input power, and use in explosive or corrosive environments. This manual is written in Chinese.
| Rated power | 300 VA |
|---|---|
| Power consumption | Less than 300 VA |
| Dimensions (W × H × D) | 350 mm × 110 mm × 430 mm |
| Weight | Approximately 15 kg |
| Input voltage and frequency | 110 V or 220 V; 47–63 Hz |
| Operating humidity | 20% to 80% RH, no condensation |
| Storage humidity | Less than 90% RH, no condensation |
| Safety standard | GB4793.1-2007 / IEC61010-1:2001 |
Turn off the instrument's power switch and unplug the power cord from the outlet; both actions must be performed.
The instrument may be used at 20% to 80% RH, with no condensation allowed.
The instrument must be connected to electrical ground, either through a grounded three-wire power outlet or by connecting the rear protective terminal to earth.