Masimo Rad-5/Rad-5v Signal Extraction Pulse Oximeter Service Manual

Service manual for the Masimo Rad-5 and Rad-5v Signal Extraction pulse oximeters: cleaning, battery replacement, performance verification and repair.


Masimo Rad-5/Rad-5v Signal Extraction Pulse Oximeter Service Manual - cover page
Brand
Masimo
Category
Health Care
Document type
Service Manual
Language
English
Pages
17
File format
PDF
File size
122 KB
Published
16 November, 2026
Updated
16 November, 2026
MD5 checksum
903C9B0DB7E001D8E69B4BC2D44879AA

About this manual

The Masimo Rad-5/Rad-5v Signal Extraction Pulse Oximeter Service Manual (UM-10026) provides the information needed to service all models of the Rad-5 class of pulse oximetry systems. It is structured to support troubleshooting to the assembly or module level and does not provide instructions for troubleshooting to the printed circuit board component level; service training provided by Masimo is required before using it.

The manual covers warnings, cautions and notes, maintenance and cleaning with 70% isopropyl alcohol, battery operation and replacement with four AA alkaline batteries, performance verification (power-on self-test, key press button test, alarm limit test on the Rad-5, display contrast test, testing with the Masimo Set Tester and battery test), repair and troubleshooting, the repair and returns procedure, and drawings and schematics.

What's inside

  • Overview
  • Warnings, Cautions and Notes
  • Maintenance
  • Cleaning
  • Battery Operation and Maintenance
  • Replacing the Batteries
  • Performance Verification
  • Power-On Self-Test
  • Key Press Button Test
  • Alarm Limit Test (Rad-5 only)

Specifications

Battery typeFour AA alkaline batteries
Cleaning solution70% isopropyl alcohol
SpO2 reading with Masimo Set Tester79% to 84%
Pulse rate reading with Masimo Set Tester55 bpm to 65 bpm
Alarm limits used for the tester checkSpO2 high OFF, low 90; pulse rate high 140, low 50

Frequently asked questions

How do I clean the Rad-5 pulse oximeter?

Turn the unit off, then wipe the display panel with a cotton swab and the outer surface with a soft cloth, both moistened with 70% isopropyl alcohol. Do not autoclave, pressure sterilize, soak or immerse the oximeter in any liquid.

What batteries does the Rad-5 use?

The Rad-5 Handheld uses four AA alkaline batteries; Masimo advises replacing them as soon as the final battery charge indicator starts flashing.

What readings should the Masimo Set Tester produce during performance verification?

Within 20 seconds the display should show numeric readings, with SpO2 between 79% and 84% and pulse rate between 55 bpm and 65 bpm.

How do I check the display contrast?

With the unit turned on, press the MODE-ENTER button three times until the display appears as 888, then use the up and down arrows to verify the contrast brightens and dims. Return the contrast to its original value afterwards.