Installation and calibration instructions for LTH ECS40T electrodeless conductivity sensors, including setup with BC9 series, HET63 and MXD70 instruments.
This LTH document explains how ECS40T electrodeless conductivity sensors are installed and calibrated. Calibration is needed only at commissioning or when the sensor or its cable is changed, and the procedure depends on the instrument the sensor is connected to, so the instructions are meant to be used together with the relevant instrument manual. In every case the loop resistor(s) must be inserted through the sensor as shown.
The document gives step-by-step display calibration for the BC9 series, menu-driven calibration for the HET63 and MXD70 instruments, and general installation precautions. It also lists guidance for the ECS42T dip sensor and for the ECS43T, ECS45T, ECS47T and ECS49T flowline, insertion and hygienic sensors, plus a list of sensor part numbers.
| Pink loop resistor value | 5 Ohm |
|---|---|
| Minimum pipe inside diameter for flowline/insertion sensors | 2.5" DN63 |
| Minimum clearance around sensor in smaller pipes | 12 mm |
| ECS42T dip sensor lengths | 600 mm (Part No 8515) and 1200 mm (Part No 8516) |
| ECS43T flowline sensor | 2" PVC, Part No 8523 |
| ECS45T flowline sensor | 2" stainless steel, Part No 8525 |
| Document issue and part number | Issue 0314, Part No. 6138 |
It only needs calibration during commissioning, or if the sensor or the cable is changed. The procedure varies with the instrument being used.
A pink 5 Ohm loop resistor is attached to the sensor through the base and out through one of the side holes.
No, the cable should not be used to suspend the sensor, and the cable or connector must not be submerged.
The cross hole should be lined up with the flow, and air must not be able to trap around the sensor.