Operator's manual for the LeCroy 3412/3412E 16-channel, 200 MHz CAMAC discriminator, covering setup, operation, theory, and specifications.
This manual covers the LeCroy Model 3412/3412E 16-channel, 200 MHz CAMAC discriminator. It provides setup and operation instructions, plus theory of operation and other functional information.
It includes installation notes on cooling and crate power, threshold and output width adjustment, veto and mask inputs, test mode, and CAMAC commands and function codes.
| Cooling exhaust air temperature | Less than 50 °C |
|---|---|
| Power requirement | Significant power from the -6 V line |
| Threshold voltage range | 0 to -10 V |
| Remote threshold DAC output | Vo = 5(X/4095) V |
| Width voltage range | About 1.4 to 11.4 V |
| Output pulse width range | Less than 5 ns to over 100 ns |
| Fast veto timing | Leading edge must precede input by at least 6 ns |
| Rear panel mask levels | ECL logic 1 true, ECL 0 complementary |
LeCroy warrants its instrument products to operate within specifications under normal use and service for one year from the date of shipment.
Fans must move sufficient air to keep exhaust air temperature below 50 °C.
A 12-bit DAC (AD7245) sets it, with output voltage Vo = 5(X/4095) V.
Its leading edge must precede the input to be vetoed by at least 6 ns.