Safety-related instructions and technical data for the LABOM Ci4 transmitter series, including SIL2 parameters, operating modes, failure rates, and…
This document provides safety-related instructions and technical specifications for the PASCAL Ci4 Type series CI4xxx transmitters manufactured by LABOM. It details the device's Safety Integrity Level (SIL) capabilities according to IEC/EN 61508, specifically covering SIL2 requirements.
| Safety Integrity Level | SIL2 |
|---|---|
| Operating Mode | Low and continuous demand mode |
| Architecture | 1oo1 |
| Device Type | Type B |
| Hardware Fault Tolerance | HFT 0 |
| Safe Failure Fraction (SFF) | >90% |
| Failure rate for safe detected failures | 2,03∙10-7 1/h |
| Failure rate for safe undetected failures | 2,37∙10-7 1/h |
| Reaction time in event of request | 220 ms |
| Reaction time in event of fault detection | 100 ms |
The safe function according to IEC/EN 61508 is the 4…20 mA current signal.
A current of < 3.6 mA (alarm current) is issued at the output until initialization completes.
An alarm current caused by critical malfunction can only be deactivated by a manual restart.