Operating instructions for Labom BX1008 high-temperature pressure gauges with contact, covering installation, electrical connection, adjustment, and…
This document provides operating instructions for the Labom BX1008 series of high-temperature pressure gauges equipped with contacts. It covers general safety notes, intended use, transportation, storage, and conformity with EU regulations including ATEX approval.
The guide details mechanical installation procedures, electrical connections, switch function identification, and inductive contact specifications. It also includes instructions for adjusting contacts, correcting the zero point, maintenance recommendations, and safe disassembly procedures.
| Permissible media temperature | 10…350 °C |
|---|---|
| Permissible ambient temperature | 10…70 °C |
| Nominal voltage (NAMUR) | 8 V= |
| Operating voltage range | 5 - 25 V |
| Current consumption (active face uncovered) | > 3 mA |
| Switching accuracy | approx. 0,5% of full scale value |
The installation, set up, service or disassembly must only be done by trained, qualified personnel using suitable equipment.
The maximum tightening torque is 60 Nm (slightly oiled).
Yes, when properly installed it is maintenance-free, though annual recalibration is recommended.