Kleinwächter EFM-022 Electrostatic Field Meter and Kit Operation Manual

The manual provides instructions for using, measuring, maintaining and calibrating the Kleinwächter EFM-022 electrostatic field meter and kit.


Kleinwächter EFM-022 Electrostatic Field Meter and Kit Operation Manual - cover page
Brand
Kleinwächter
Category
Other device
Document type
User Manual
Language
Chinese
Pages
15
File format
PDF
File size
1.1 MB
Published
06 October, 2026
Updated
06 October, 2026
Model
EFM-022
MD5 checksum
768AA903594A04C641A41C4C59E358CB

About this manual

The Kleinwächter EFM-022 is a portable electrostatic field meter and kit that can measure surface static voltage, electrostatic field strength, ion‑balance of static eliminators and human walking static voltage. It features a rotating‑blade probe, EMV‑shielded housing and interchangeable accessories such as CPS‑022 high‑voltage generator and VMS accessories.

The operation manual explains how to set up the instrument, select measurement modes (static voltage, E‑Field, CPS‑Mode, Max‑Value), perform measurements, perform routine maintenance, calibrate the device and understand warranty responsibilities. Safety warnings, grounding requirements and battery information are also included. This manual is written in Chinese.

What's inside

  • Order Numbers
  • Product Description
  • Instrument Functions
  • Surface Static Voltage Measurement
  • Electrostatic Field Strength Measurement
  • CPS Mode Test
  • Max Value Mode Test
  • Maintenance, Calibration and Warranty

Specifications

Static voltage range (1 cm distance)0–10 kV
Static voltage range (20 cm distance)0–200 kV
CPS‑Mode voltage range100–1000 V (resolution 1 V)
Measurement accuracy±5 %
Power supply9 V battery (ordinary or rechargeable)
Probe input impedance>10^16 Ω
Probe input capacitance<1 pF
Dimensions and weight70 × 122 × 26 mm, 130 g

Frequently asked questions

How is the EFM-022 powered and what are the low‑battery warnings?

The instrument runs on a 9 V battery (ordinary or rechargeable). When the voltage drops below 7.6 V the screen shows “Low Battery”, and below 7.1 V it displays “auto off” and shuts down after 2 seconds.

How do you change measurement modes on the EFM-022?

Long‑press the function key until “Change Mode” appears, then press the key once to cycle to the next mode. The instrument will start measuring in the selected mode after the indicator disappears.