This document details the technical specifications for Keithley’s Model 2635A and 2636A System SourceMeters. It covers voltage and current source capabilities, measurement accuracy, noise levels, and operating conditions. Specifications are valid for both Normal and High Capacitance Modes and apply at 23°C ± 5°C, with a one-year calibration period. The devices support four-quadrant source or sink operation and include pulse mode capabilities for specific current limits and widths.
| Voltage Source Accuracy (1 Year) | ± (% rdg. + volts) |
|---|---|
| Current Source Accuracy (1 Year) | ± (% rdg. + amps) |
| Maximum Output Power | 30.3 W per channel maximum |
| Temperature Coefficient | ± (0.15 × accuracy specification)/°C |
| Voltage Output Headroom (200 V Range) | Maximum output voltage = 202.3 V – total voltage drop across source leads |
| Pulse Width Programming Resolution | 1 µs |
| Pulse Width Programming Accuracy | ± 5 µs |
Maximum output power is 30.3 W per channel.
Voltage source accuracy is ± (% rdg. + volts).
Temperature coefficient is ± (0.15 × accuracy specification)/°C.