Interscan GF1900 Sulfuryl Fluoride Analyzer user manual covering operation, calibration, maintenance, troubleshooting, parts, and return authorization.
The Interscan GF1900 Sulfuryl Fluoride gas fumigant analyzer is designed for use as a clearance monitoring device to determine safe re-occupation of a structure after aeration. The manual describes its controls, operation, calibration, maintenance, troubleshooting, accessories, warranty, diagrams, parts list, and return authorization process.
Key cautions include the furnace's limited life, with a mean time before failure of 1000 hours and reduced life from exposure above 50 ppm, and the need to avoid use where gasoline, sulfur compounds, or highly flammable vapors are present. Operation and calibration instructions cover power connection, zeroing, battery use, recharge time, tubing length, and 5 ppm calibration gas handling.
| Furnace operating temperature | 900 degrees C |
|---|---|
| Furnace mean time before failure | 1000 hours |
| Sulfuryl fluoride level that can reduce furnace life | Above 50 ppm |
| Recommended sample tubing | Teflon, 1/4 in OD x 1/8 in ID |
| Maximum tubing length with factory calibration | 12 inches |
| Tubing length requiring calibration with attached length | 12 to 48 inches |
| Battery operating time | Approximately 70 minutes |
| Battery recharge time | 8 to 10 hours |
| Low battery indication threshold | 19.0 V DC |
| Calibration gas concentration | 5 ppm sulfuryl fluoride |
| Calibration repeatability | Within ±2 divisions |
| New or replacement sensor stabilization time | Approximately 24 hours |
It is expressly designed for use as a clearance monitoring device to determine safe re-occupation of a structure after aeration.
Calibration should be done every 3 months or more frequently; some companies require monthly calibration.
The instrument is calibrated using a known concentration of sulfuryl fluoride gas at 5 ppm.
Approximately 70 minutes of operating time is available with a fully charged battery pack.