Interscan GF1900 Sulfuryl Fluoride Gas Fumigant Analyzer User Manual

Interscan GF1900 Sulfuryl Fluoride Analyzer user manual covering operation, calibration, maintenance, troubleshooting, parts, and return authorization.


Interscan GF1900 Sulfuryl Fluoride Gas Fumigant Analyzer User Manual - cover page
Brand
Interscan
Category
Analyzers
Document type
User Manual
Language
English
Pages
23
File format
PDF
File size
1 MB
Published
13 November, 2026
Updated
13 November, 2026
MD5 checksum
1AF5E47424B2A7A02C9107A193C7549C

About this manual

The Interscan GF1900 Sulfuryl Fluoride gas fumigant analyzer is designed for use as a clearance monitoring device to determine safe re-occupation of a structure after aeration. The manual describes its controls, operation, calibration, maintenance, troubleshooting, accessories, warranty, diagrams, parts list, and return authorization process.

Key cautions include the furnace's limited life, with a mean time before failure of 1000 hours and reduced life from exposure above 50 ppm, and the need to avoid use where gasoline, sulfur compounds, or highly flammable vapors are present. Operation and calibration instructions cover power connection, zeroing, battery use, recharge time, tubing length, and 5 ppm calibration gas handling.

What's inside

  • Introduction
  • Cautions
  • Operation
  • Calibration
  • Maintenance
  • Trouble Shooting
  • Accessories
  • Warranty
  • Diagrams, Schematics, and Parts
  • Return Authorization

Specifications

Furnace operating temperature900 degrees C
Furnace mean time before failure1000 hours
Sulfuryl fluoride level that can reduce furnace lifeAbove 50 ppm
Recommended sample tubingTeflon, 1/4 in OD x 1/8 in ID
Maximum tubing length with factory calibration12 inches
Tubing length requiring calibration with attached length12 to 48 inches
Battery operating timeApproximately 70 minutes
Battery recharge time8 to 10 hours
Low battery indication threshold19.0 V DC
Calibration gas concentration5 ppm sulfuryl fluoride
Calibration repeatabilityWithin ±2 divisions
New or replacement sensor stabilization timeApproximately 24 hours

Frequently asked questions

What is the Interscan GF1900 used for?

It is expressly designed for use as a clearance monitoring device to determine safe re-occupation of a structure after aeration.

How often should the GF1900 be calibrated?

Calibration should be done every 3 months or more frequently; some companies require monthly calibration.

What concentration of sulfuryl fluoride is used for calibration?

The instrument is calibrated using a known concentration of sulfuryl fluoride gas at 5 ppm.

How long does a fully charged battery pack operate?

Approximately 70 minutes of operating time is available with a fully charged battery pack.