The Instech Model 100 Syringe Infusion Pump Operations Manual provides setup, programming, safety, and maintenance instructions for the pump.
This manual explains how to install, program, and operate the Instech Model 100 syringe infusion pump. It includes safety warnings, detailed descriptions of the pump’s front and back panels, menu navigation, and step‑by‑step procedures for loading syringes, setting infusion modes, and delivering medication. Key sections cover general specifications, alarm and alert handling, battery operation, IV‑pole mounting, cleaning, and service warranty information, plus appendices with flow‑rate tables, syringe size data, and occlusion force values.
| Overall Size | 9.5" wide × 3.95" high × 5.9" deep |
|---|---|
| Weight | 4.99 Pounds |
| Pump Accuracy | ±3% |
| Power Source | AC Power or DC |
| Battery Capacity | 8‑hour charge provides ~12 h operation at 5 mL/hr (60 mL syringe) |
| Relative Humidity | 15 % to 95 % non‑condensing |
| Ambient Pressure | 70 kPa to 106 kPa (10.2 psi to 15.4 psi) |
The battery can be fully recharged in no longer than 8 hours and will operate the pump for about 12 hours at 5 mL/hr with a 60 mL syringe.
Supported syringes include Becton Dickinson (B‑D), Monoject, Terumo, B‑D Glass, and custom XXX syringes with the listed sizes.