The Galvanic Applied Sciences Model 902 is a lead acetate tape analyzer for measuring H2S and total sulphur in gaseous streams, featuring up to eight stream…
The Galvanic Applied Sciences Model 902 H2S / Total Sulphur Analyzer measures hydrogen sulfide (H2S) in gaseous streams using the lead acetate tape method. It is designed for applications such as refineries, gas processing facilities, and pipeline custody transfer. The unit can also measure total sulphur when equipped with an optional furnace.
| Ranges | 0-100 ppb to 0-50 ppm; 50 ppm to 100% with dilution |
|---|---|
| Response time | 3 minutes to 90% |
| Streams | 1 standard plus calibration; Up to 8 optional |
| Linearity | ±1% of full scale or 0-30ppm ±5 for all other ranges |
| Power consumption | 10 Watts @ 24 VDC or 120 VAC 60 Hz |
| Tape Life | 4-8 weeks, depending on application |
| Outputs | 2 isolated 4–20 ma; 6 relay contact (SPDT - 5 amps, 250 VAC max) |
| Microprocessor | 87 C530 @ 24 MHz |
The analyzer uses the lead acetate tape method, where H2S reacts with lead acetate to form a brown stain.
Yes, it can measure total sulphur if equipped with an optional hydrogenation reaction furnace.
Calibration is only required every 3-4 months because the system self-zeros at every tape advance.
The unit has ATEX EEx d [ia] ia IIB+H2 T3(TA5) and Russian Ex Proof Certification.