G.R.A.S. 46BE 1/4" CCP free-field standard microphone set: 4 Hz to 80 kHz frequency range, 35 dBA to 160 dB dynamic range, 4 mV/Pa sensitivity, TEDS compliant.
The G.R.A.S. 46BE is a 1/4" Constant Current Power (CCP) free-field standard microphone set for measuring high sound pressure levels at high frequencies, with a frequency range of 4 Hz to 80 kHz and a dynamic range of 35 dBA to 160 dB. It consists of a G.R.A.S. 40BE 1/4" prepolarized free-field microphone cartridge and a G.R.A.S. 26CB 1/4" CCP standard preamplifier, calibrated together as one unit, and it is terminated with a 10/32 UNC Microdot female connector. The set is IEEE 1451.4 TEDS v. 1.0 compliant; the equivalent traditional 200 V LEMO type is the G.R.A.S. 46BF.
Because the cartridge and preamplifier are calibrated as one unit, only one sensitivity value has to be accounted for. The 46BE is optimized for 0° incidence free-field measurements, but the small 1/4" body and diaphragm also allow pressure and random-incidence measurements up to 20 kHz. The document describes the design, compatibility requirements, system verification, service, calibration and warranty, and lists full specifications, typical frequency response, dimensions, ordering information and optional items. G.R.A.S. gives a 5-year warranty against defective materials and workmanship and recommends recalibration at least once a year.
| Frequency range (±1 dB) | 10 Hz to 40 kHz |
|---|---|
| Frequency range (±2 dB) | 4 Hz to 80 kHz |
| Frequency range (±3 dB) | 4 Hz to 100 kHz |
| Dynamic range lower limit | 35 dB(A) |
| Dynamic range upper limit | 160 dB |
| Set sensitivity @ 250 Hz | 4 mV/Pa (-48 dB re 1 V/Pa, ±2 dB) |
| Output impedance | < 50 Ω |
| Power supply | 2 to 20 mA |
| DC bias voltage (typ.) | 12 V |
| Operating temperature range | -30 to 70 °C / -22 to 158 °F |
| Humidity range | 0 to 100 % RH, non condensing |
| TEDS | UTID 27, IEEE 1451.4 v. 1.0 |
| Connector type | Microdot 10/32 |
| Compliance | IEC 61094-4 WS3F; CE/RoHS compliant/WEEE registered |
| Weight | 8 g / 0.282 oz |
It requires a constant current input module that can deliver 4 mA and 24 V unloaded CCP voltage supply. A lower current or voltage supply reduces the cable-driving capability or affects the upper dynamic range.
Yes, the 46BE is IEEE 1451.4 TEDS v. 1.0 compliant, so data such as properties and calibration data can be read and written when the measurement platform supports Transducer Electronic Data Sheets.
The equivalent traditional 200 V LEMO type is the G.R.A.S. 46BF.
G.R.A.S. recommends recalibrating the microphone at least once a year, depending on use, measurement environment and internal quality control programs.