Specification for the FRIWO FW7405M/24 medical AC/DC adapter: 100-240 V AC input, 24 V DC 2.2 A output, housing, leads, packaging and safety data.
This is FRIWO's delivery specification (Liefervorschrift / Specification) for type FW7405M/24, part number 1825898, drawing number 15.2217.500-00, document number P000417848, index ⓔ dated 2012/07/10. The nine-page German/English document covers the housing, housing labelling, leads, packaging, general and electrical test conditions, safety details and the CE declaration of conformity.
According to the packaging print, the unit is an AC/DC adapter with an input of 100-240 V AC and an output of 24 V DC / 2.2 A. The document lists nominal input and output data measured at +20 °C after 15 minutes of operation, an application range for medical applications and a safety build according to IEC60601-1 / ES60601-1 with protection class II. This manual is written in German.
| Nominal input voltage | 100-240V AC ±10% |
|---|---|
| Nominal input frequency | 50-60Hz |
| Nominal input current | 1,100 - 0,500Arms @ max load |
| Nominal output voltage | 24V DC +4% / -4% |
| Output ripple | ≤ 120mVpp |
| Nominal output current | 0 - 2200mA |
| Stand-by power consumption at 230V AC | ≤ 0,75W |
| Protection class | II |
| Safety standard | IEC60601-1, ES60601-1 |
| Leakage current | ≤ 10µA |
| High-voltage test | ≥ 4kVac |
| Ambient temperature range | 0°C to +40°C |
| Storage temperature | -40°C - +70°C / 10 to 95 rel. hum. |
| Weight per unit | 250 g |
| Output lead length | 1830 mm |
| Housing material | PC / ABS V0 125°C |
| Units per master carton | 40 |
The nominal output voltage is 24 V DC with +4% / -4% tolerance and the nominal output current is 0 to 2200 mA.
It accepts a nominal input voltage of 100-240 V AC ±10% at a nominal input frequency of 50-60 Hz.
The safety build follows IEC60601-1 and ES60601-1, protection class II, for medical applications.
Faultless function must be guaranteed from 0°C to +40°C at 90% relative humidity without condensation.