Fox Thermal Instruments FT3 Data Logger User Manual

Operation and configuration guide for the Fox Thermal Instruments FT3 data logger option, covering menu trees, setup and Modbus commands.


Fox Thermal Instruments FT3 Data Logger User Manual - cover page
Brand
Fox Thermal Instruments
Category
Mass Flowmeter
Document type
User Manual
Language
English
Pages
18
File format
PDF
File size
1 MB
Published
17 November, 2026
Updated
17 November, 2026
MD5 checksum
27F8ACADF6DC52D8F047E25DE53CCB2C

About this manual

This manual describes the operation and configuration of the data logger option available on the Fox Thermal Instruments Model FT3 Thermal Gas Mass Flow Meter and Temperature Transmitter. The logger records interval totals based on the selected flow units at a customer-specified rate and keeps the most recent 31 records with a date and time stamp.

The document covers the front-panel menu trees, setup of the data logger, start/sync time and real time clock, viewing and clearing records, the Modbus commands that support reading and clearing the log, and a glossary of terms. It is marked 105774 Rev. B and applies to FT3 firmware version 3.09 and later.

What's inside

  • Introduction
  • Menu Trees
  • Data Logger Settings
  • Data Logger Display
  • Setup
  • Clearing the Data Log
  • Setting the Start/Sync Time
  • Setting the Real Time Clock
  • Operation: Configure
  • MODBUS Data Log Support

Specifications

Records stored31 separate 24-hour daily totals with date and time stamp; oldest data overwritten
Recording intervalCustomer selectable from 1 second to 1 month
Default start timeMidnight
Data accessRS485 Modbus and the Engineering Display
Manual revision compatibilityRev B: FT3 firmware v 3.09 and later
Clear log Modbus registerPreset Register, Modbus Address 40213, data 0x57

Frequently asked questions

How many records can the FT3 data logger store?

It supports 31 records. When the number of samples exceeds 31, the old data is overwritten so only the most recent 31 records are kept.

How do I view data logger records from the FT3 front panel?

Press the F1 and F4 keys at the same time. Use F2 to move to the next record, F1 to move back, and F4 to exit to normal mode.

What is the default start/sync time for the 24-hour total?

The default is midnight (day, hour, minute and second all zero); only the hour, minute and second are used for the trigger.

How is the data log cleared over Modbus?

Command 06 (Preset Register) at Modbus address 40213 with data 0x57 is used to clear all records in the log.

Download PDF (1 MB)