This manual describes the installation and use of the Fox Thermal Instruments Model 10A temperature sensor.
The Model 10A is an advanced, microprocessor-based Thermal Mass Flowmeter designed to measure gas flow directly in mass units (SCFM, NM3/Hr, Kg/Hr). It operates using a balanced bridge circuit with heated sensors to maintain high accuracy and repeatability independent of temperature and pressure changes. The device features a field-programmable display and supports various flowmeter styles, including insertion and flow body styles.
This document covers the complete lifecycle of the instrument, including theory of operation, detailed installation procedures for different mounting styles, display and menu programming options, wiring configurations for power and output signals, preventative maintenance steps, and troubleshooting guides. It also specifies physical dimensions, performance capabilities, and environmental ratings for both the sensor and optional remote enclosures.
| Accuracy | +/- 1.0% of reading + 0.2% of full scale |
|---|---|
| Response Time | 0.9 seconds (One time constant) |
| Repeatability | +/- 0.2% of full scale |
| Calibration | NIST traceable |
| Sensor Temperature Range | -40 to 250°F (-40 to 121°C) |
| Input Power (Explosion Proof) | 24 VDC +/- 10% 0.75 amp |
| Output Signal | 4-20mA, isolated |
| Alarm Output | Alarm relay; dry contact; 50 VAC, 30 VDC, 3 amps |
| Remote Cabling | 2 conductor, twisted shielded, 18 AWG, 500 ft. max |
The device supports SCFM, SCFH, NM3H, NLPM, KG/HR, LBS/H, LBS/M, NMPS, and SFPM.
No, the Model 10A measures mass flow directly, eliminating inaccuracies associated with volumetric flow devices that depend on temperature and pressure.
The maximum gas pressure rating for the NPT connection is 500 psig (34.5 barg).
The sensor wetted materials are 316 Stainless Steel, with Hastelloy C-276 available as an option.