User manual for the Fluke VT650/VT900A gas flow analyzer, covering setup, measurements, maintenance, calibration, and technical specifications.
The Fluke VT650/VT900A Gas Flow Analyzer is a portable gas flow analyzer and ventilator tester for measuring pressure, flow, volume, oxygen concentration, and gas temperature. It is intended for testing ventilators and medical gas flow devices in laboratory environments, not for direct patient use.
The manual covers unpacking, connections, test setup, operation, measurement menus, maintenance, calibration, and technical specifications. The VT900A adds ultra-low flow and ultra-low pressure measurement ranges and an external trigger input. This manual is written in Chinese.
| Ultra-low flow range (VT900A) | ±750 ml/min |
|---|---|
| Ultra-low pressure range (VT900A) | 0 mbar to 10 mbar |
| Battery operation time | Up to 8 hours |
| Airway flow sampling frequency | ≥ 200 Hz |
| Airway flow rise time | <10 ms |
| Airway flow volume range | 100 l |
| Temperature range | 0 °C to 50 °C |
| Humidity range | 0% to 100% relative humidity |
| Oxygen concentration accuracy | ±1% (VT900A), ±2% (VT650) |
| General safety standard | IEC 61010-1: Pollution Degree 2 |
| Lithium-ion battery standard | IEC 62133 |
| Calibration interval | At least every 12 months |
| Warranty period | One year from original purchase |
It is a portable gas flow analyzer and ventilator tester used to test ventilators and medical gas flow devices for preventive maintenance, repair inspection, and routine inspection.
No. It is intended for laboratory environments outside patient care areas and is not for direct patient use or testing devices connected to patients.
The VT900A can measure ultra-low flow of ±750 ml/min and ultra-low pressure from 0 mbar to 10 mbar, and it includes an external trigger input.
Fluke Biomedical recommends calibration at least every 12 months, performed by qualified personnel.
The battery provides up to 8 hours of operation.