Fluke Networks DSP-100/DSP-2000 Cable Test Tool Users Manual supplement with calibration notes, return loss specs, wire maps and LinkWare details.
This document is Issue 5 of the DSP-100/DSP-2000 Users Manual Supplement from Fluke Networks, dated 10/02. It lists changes and corrections to the Users Manual, including notes on self-calibration, a replacement Return Loss specification, and revised Wire Map Displays.
Further changes cover traceable calibration once a year, a warning not to use a NiMH battery pack in place of the NiCad pack, and the inclusion of LinkWare Cable Test Management software with the DSP-2000.
| Return loss range | 0 to 30 dB |
|---|---|
| Return loss accuracy | ± 3 dB over the following ranges |
| Return loss accuracy range (5 to 10 MHz) | 0 to 18 dB |
| Self-calibration ambient temperature | 10°C to 40°C (50°F to 104°F) |
| Self-calibration warm-up time | 1 minute |
The DSP-2000 includes LinkWare Cable Test Management software instead of CableManager or DSP-LINK software.
It requires a traceable calibration once a year.
No. The test tool will not charge properly with a NiMH battery pack.
Allow a 1 minute warm-up time before performing self-calibration.