FEI 2LE Two Lens Electron Column User's Guide

FEI 2LE Two Lens Electron Column User's Guide covering installation, operation, maintenance, and troubleshooting for the UHV SEM column.


FEI 2LE Two Lens Electron Column User's Guide - cover page
Brand
FEI
Category
Other Electronics
Document type
User Manual
Language
English
Pages
110
File format
PDF
File size
5.4 MB
Published
08 November, 2026
Updated
08 November, 2026
MD5 checksum
1E5EC70C9BD35501070DF3F47D26619F

About this manual

FEI's Two Lens Electron (2LE) Column is designed for use in an ultra-high vacuum (UHV) environment and provides bolt-on scanning electron microscope (SEM) capabilities to an existing system. It uses an FEI thermal field emitter with a single-crystal, zirconium oxide coated tungsten emitter and an automatically variable aperture.

The column resolves features as small as 20 nm under typical imaging conditions and the guide covers installation, operation, user maintenance, and troubleshooting. Safety, handling, vacuum, and high-voltage precautions are addressed throughout the manual.

What's inside

  • System Overview
  • Safety and Handling
  • Hardware and Theory
  • Scanning System Installation
  • Column Operation
  • User Maintenance
  • Troubleshooting
  • Technical Articles
  • Glossary
  • Index

Specifications

Operating environmentUltra-high vacuum (UHV)
SEM capabilityBolt-on scanning electron microscope (SEM) capabilities
EmitterSingle-crystal, zirconium oxide coated tungsten (ZrO/W) with <100> orientation
ResolutionResolves features as small as 20 nm
Typical imaging conditions1 nA probe current, 25 mm working distance, 25 kV beam voltage
Working distanceRelatively large (> 25 mm)
Electron beam energy spread0.5 eV
Maximum operating pressure5 x 10^-9 torr
Optical system magnificationUnit magnification for a 6 nA beam

Frequently asked questions

What environment is the FEI 2LE column designed for?

It is designed for use in an ultra-high vacuum (UHV) environment.

What SEM capability does the 2LE column provide?

It provides bolt-on scanning electron microscope (SEM) capabilities to an existing system.

What is the maximum pressure at which the column can be operated?

The column must never be operated with pressure greater than 5 x 10^-9 torr.

What emitter does the column use?

It uses a single-crystal, zirconium oxide coated tungsten (ZrO/W) emitter with <100> orientation.

Download PDF (5.4 MB)