User manual for the EDI TT-1 Transducer Test Set, a device to test transducer performance, identify leaks, and determine operating frequency.
The TT-1 Transducer Test Set is a tool designed to identify leaky transducers, measure their activity, and determine their operating frequency. It uses an R-C bridge and a sensitive phase detector to analyze transducer performance. The device can also provide a 'signature' of a newly installed transducer for future reference.
Operation involves connecting the transducer, selecting capacitance ranges, nulling out cable and element capacitance, and sweeping through frequency ranges to detect phase transitions. The highest or broadest transition point typically indicates the operating frequency. The device can also be used to record resonance points for later comparison and to test transducer sensitivity by applying hand pressure.
It identifies leaky transducers, indicates the degree of activity of the element, and aids in identifying the operating frequency of the transducer.
It uses an R-C bridge combined with a sensitive phase detector to measure and analyze transducer performance.
Sweep through the frequency range and note the points at which the phase indicator changes; the operating frequency is usually indicated by the highest or broadest transition point.